6
HA0200T Rev 11 Jun 2022
Chapter 3
5) Connect the foot pump to the mount on one of the legs and inflate to about 40 p.s.i. – see Fig. 3.5.
6) Inflate the leg diagonally opposite to 40 p.s.i.
7) Repeat steps 5) and 6) for the two remaining legs.
8) Populate the breadboard.
9) Further inflate the mounts until the breadboard is level and a gap of 12mm between the underside of the breadboard and
the top of the mount is obtained – see Fig. 3.6.
10)Fit the dust caps.
Fig. 3.5 Inflating passive anti-vibration mounts
Caution
During item (8), the total load including the breadboard must not exceed the following figures:
Standard mounts – 260kg (572 lb)
Heavy duty mounts – 720kg (1584 lb)
Caution
During item 9):
It may be necessary to inflate each mount to a different pressure due to the loading of the breadboard
Do not exceed 72 p.s.i.