Optical Spectrum Analyzers
Chapter 4: Description
Page 5
STN053070-D02
The maximum physical movement of the OSA20xC platform is 10 mm from the ZPD, which translates to a
maximum OPD of 40 mm and a spectral resolution of 7.5 GHz or 0.25 cm
-1
. The Redstone OSA305 has a
80 mm stroke length capacity, and thus a maximum OPD of 160 mm and a spectral resolution of 1.9 GHz or
0.063 cm
-1
. The wavelength resolution is dependent on the wavelength of light being measured. For more details
see Section 4.8 Resolution and Sensitivity. In this context, the spectral resolution is defined according to the
Rayleigh criterion and is the minimum separation required between two spectral features in order to resolve
them as two separate lines. Note that when using the Wavelength Meter or Peak Track tools, the number of
significant figures can far exceed the Rayleigh spectral resolution as those tools use several data points and
curve fitting techniques.
The Thorlabs OSAs utilize a built-in, actively stabilized reference laser to interferometrically record the variation
of the optical path length. This reference laser is inserted into the interferometer and closely follows the same
path traversed by the unknown input light field. To reduce the effects of water absorption on the acquired
spectrum, our OSAs have two 1/4" inner-diameter quick-connection hose fittings on the back panel, through
which the interferometer can be purged with dry air or nitrogen, as detailed in Chapter 6.
Figure 1. Optical Schematic of the Thorlabs OSA20xC Detailing the Dual Retroreflector Design
Содержание Redstone OSA305
Страница 1: ...Redstone OSA305 OSA201C OSA202C OSA203C OSA205C OSA207C Optical Spectrum Analyzers Manual ...
Страница 43: ...Optical Spectrum Analyzers Chapter 12 Mechanical Drawings Page 39 STN053070 D02 12 1 2 Front and Back Panels ...
Страница 45: ...Optical Spectrum Analyzers Chapter 12 Mechanical Drawings Page 41 STN053070 D02 12 2 2 Front and Back Panels ...
Страница 52: ...www thorlabs com ...