26
PM100USB
7.1.2.3.3 CALibration subsystem commands
Command
Description
CALibration
Path to CALibration subsystem. (SCPI Vol.2 §5)
:STRing?
Returns a human readable calibration string. This is a
query only command. The response is formatted as string
response data.
7.1.2.3.4 SENSe subsystem commands
Command
Description
[SENSe]
Path to SENSe subsystem. (SCPI Vol.2 §18)
AVERage
[:COUNt] <value
Sets the averaging rate (1 sample takes approx. 3ms)
[:COUNt]?
Queries the averaging rate
CORRection
[:LOSS[:INPut[:MAGNitude]]]
{MINimum|MAXimum|DEFault|
<numeric_value>}
Sets a user attenuation factor in dB
[:LOSS[:INPut[:MAGNitude]]]?
[{MINimum|MAXimum|DEFault}]
Queries the user attenuation factor
COLLect
ZERO
[:INITiate]
Performs zero adjustment routine
ABORt
Aborts zero adjustment routine
STATe?
Queries the zero adjustment routine state
MAGNitude?
Queries the zero value
BEAMdiameter {MINimum|
MAXimum|DEFault|
<numeric_value>[mm]}
Sets the beam diameter in mm
BEAMdiameter? [{MINimum|
MAXimum|DEFault}]
Queries the beam diameter
WAVelength {MINimum|
MAXimum|
<numeric_value>[nm]}
Sets the operation wavelength in nm
WAVelength? [{MINimum|
MAXimum}]
Queries the operation wavelength
POWer
[:PDIOde]
Sets the photodiode response value in A/W
[:RESPonse] {MINimum|
MAXimum|DEFault|
<numeric_value>[A]}
[:RESPonse]?
[{MINimum|MAXimum|
DEFault}]
Queries the photodiode response value
:THERmopile