background image

© 2020 Thorlabs

6 Measurement Considerations

15

The selection of the best suitable pyroelectric energy sensor must be made with respect to the
following parameters:

Energy to Measure 

The  maximum  allowed  energy  is  limited  by  the  absorption  layer  (excessive  energy  causes
mechanical  damages  on  the  layer).  Here,  the  specified  maximum  temperature  level  must  not
be  exceeded.  It  is  important  to  pay  attention  to  the  maximum  energy  density,  to  the  average
power, and to the maximum power density during a pulse.

On the lower end, the detection is limited by the resolution and the minimum 

trigger level

, as

well as by the noise level.

Beam Diameter

The active area of the detector should be selected in such that it has a slightly larger diameter
than the incident beam. The beam should not  necessarily  cover  most  of  the  sensor  area,  but
the maximum allowed energy density must not be exceeded.

Repetition Rate

The maximum pulse repetition rate depends on the combination of the internal capacitance of
the detector and the load resistor. The display of a power meter unit has an input impedance of
1 MΩ (similar to the typical input resistor of an oscilloscope).

Pulse Length

Energy sensors can detect and measure pulses with a duration from sub-nanosecond range to
approximately 2 ms. The maximum pulse duration depends on

·

the electrical time constant of the sensor, given by detector capacity and load resistance

·

the thermal time constant of the sensor. 

Usually the latter is the more significant.

Wavelength

The  sensors  are  typically  calibrated  at  one  wavelength.  A  sensor  specific  correction  curve  is
stored in the memory of the DE-9 connector for other wavelengths. The wavelength dependent
response  is  nearly  linear  flat  for  black  coating  over  a  wavelength  range  from  185  nm  to  >  25
µm; the ceramic coating is also suitable for this wavelength range, but is not flat over the entire
wavelength range.

6.2   Reducing Noise for High Accuracy Measurements

Noise from grounding, cable capacitance, temperature effects, stray and ambient light, and de-
tector  noise  are  interfering  with  the  measurement.  The  impact  of  noise  is  high  for  measure-
ments of low optical power signal. Below are some recommendations on how to reduce inter-
ference to a minimum:

·

The housing of power sensors is connected to the digital ground of the power meter and
should be linked to earth ground (e.g. via post mounting). 

·

Energy sensors should be mounted as isolated devices, using the supplied isolation a-
dapters, because the housing is connected to the power meter analog ground.

·

The sensor cable conducts very small current or voltage signals. The cable capacitance
induces disturbances when the cable is moved. Whenever very small power or energy
levels are measured, the cable should be fixed in position. 

·

The bandwidth should be set to Low , when using photodiode sensors. 

·

The acceleration circuit should be switched off when using thermal sensors.

·

Choose a suitable sensor: The detector noise is lowest with Si or InGaAs sensors.

12

Содержание PM100USB

Страница 1: ...Optical Power Meter PM100USB Operation Manual 2020...

Страница 2: ...Version Date 1 5 13 Mar 2020 Copyright 2020 Thorlabs...

Страница 3: ...ermal Sensors 11 5 3 Pyroelectric Energy Sensors 12 5 3 1 Auto Range 12 5 3 2 Trigger Level 12 5 3 3 Average Power Measurement 12 5 3 4 Density Calculation 12 5 3 5 Custom Pyroelectric Sensors 6 Measu...

Страница 4: ...32 32 9 1 Technical Data 35 9 2 Pin Assignment of the Sensor Connector 36 9 3 Tutorial 36 9 3 1 Definitions and Explanations 36 9 3 1 1 Power Meter Specifications 37 9 3 1 2 Sensor Specifications 39...

Страница 5: ...nternational partners are looking forward to hearing from you Thorlabs GmbH Warning Sections marked by this symbol explain dangers that might result in personal injury or death Always read the associa...

Страница 6: ...via a fast USB interface which makes it easy to integrate the instrument in test and measurement systems The OPM software including the instrument drivers is available for download from the Thorlabs...

Страница 7: ...o be damaged keep it until you have inspected the contents for completeness and tested the PM100USB mechanically and electrically Verify that you have received the following items within the package 2...

Страница 8: ...ready for operation 4 LED USB connection to remote steering device PC or tablet established 3 2 Connecting a Power Sensor The PM100USB supports all Thorlabs C Series photodiode energy and thermal sens...

Страница 9: ...s Sensors without a DE 9 connector The PM100USB will NOT automatically recognize sensors lacking a DE 9 connector as for the case of the A and B series Please contact Thorlabs for an upgrade of old se...

Страница 10: ...can be found in the software manual downloaded here Note Do not connect the PM100USB prior to software installation Note The PM100USB can also be run using custom made software For this please refer t...

Страница 11: ...l response data are stored in the non volatile memory of the DE 9 connector and are automatically downloaded by the PM100USB and utility software upon plug ging to the power meter Measurement Data Whe...

Страница 12: ...is compatible with the Thorlabs C Series of thermal power sensors S3xxC S4xxC S175C equipped with a DE 9 connector The sensor is ready to operate only few seconds after plugging it to the DE 9 connect...

Страница 13: ...ty software Optical Power Monitor will automatically recognize the presence of an adapter and show Sensor Type Photodiode Adapter in the GUI 2 Calibration Data Be aware that spectral response or other...

Страница 14: ...et to exclude unwanted low power pulses from the analysis When an in coming pulse exceeds the set trigger level the peak detector circuit gets armed and is waiting until the pulse peak is reached Afte...

Страница 15: ...onitor will automatically recognize the presence of an adapter and show Sensor Type Photodiode Adapter in the GUI 2 Calibration Data Be aware that spectral response of your custom pyroelectric sensor...

Страница 16: ...laser power This heats up the absorber and the heat flow through thermocouple elements between absorbing area and sensor heat sink generates a small voltage Thorlabs C Series thermal power sensors S3...

Страница 17: ...esistance the thermal time constant of the sensor Usually the latter is the more significant Wavelength The sensors are typically calibrated at one wavelength A sensor specific correction curve is sto...

Страница 18: ...egrate the power incident on the active area of the sensor Therefore the time resolution of the thermal sensor is limited 6 4 Line Width of Light Sources The line width of light sources can be neglect...

Страница 19: ...power or energy density for the beam diameter inserted in the software For high power or high energy beams a good efficiency can be reached by selecting a sensor with a detector size about 20 30 larg...

Страница 20: ...instrument driver PM100D NI VISA to use the SCPI commands by using the Power Meter Driver Switcher that is installed with the Optical Power Monitor software package Please ensure to also install NI V...

Страница 21: ...e enclosed in square brackets The brackets indicate that the parameter is optional and can be omitted The brackets are not sent with the command string In this example DC can be omitted so the command...

Страница 22: ...ndard defines a set of common commands that perform functions like reset self test and status operations Common commands always begin with an asterisk are four to five characters in length and may inc...

Страница 23: ...the instrument will always return 0 or 1 The following command uses a boolean parameter CURRent RANGe AUTO OFF 0 ON 1 String Parameters String parameters can contain virtually any set of ASCII charac...

Страница 24: ...e query Read the Standard Event Enable Register ESR Event status register query Read and clear the Standard Event Register IDN Identification query Read the unit s identification string OPC Operation...

Страница 25: ...completed the instrument returns to the idle state at which time all pending commands including OPC and or OPC are executed After the last pending com mand is executed the OPC bit and or an ASCII 1 i...

Страница 26: ...uery information about the connected sensor This is a query only command The response consists of the following fields name sn cal_msg type subtype flags name sn cal_msg type subtype flags Sensor name...

Страница 27: ...the event register CONDition Read the condition register PTRansition value Program the positive transition filter PTRansition Read the positive transition filter NTRansition value Program the negativ...

Страница 28: ...er attenuation factor in dB LOSS INPut MAGNitude MINimum MAXimum DEFault Queries the user attenuation factor COLLect ZERO INITiate Performs zero adjustment routine ABORt Aborts zero adjustment routine...

Страница 29: ...n state UPPer MINmum MAXimum numeric_valuje A Sets the current range in A UPPer MINimum MAXimum Queries the current range REFerence MINimum MAXimum DEFault numeric_value A Sets a delta reference value...

Страница 30: ...T W DBM Sets the power unit W or dBm UNIT Queries the power unit VOLTage DC RANGe AUTO OFF 0 ON 1 Switches the auto ranging function on and off AUTO Queries the auto ranging function state UPPer MINmu...

Страница 31: ...TATe OFF 0 ON 1 STATe AUTO OFF 0 ON 1 TAU MINimum MAXimum DEFault numeric_value s TAU MINimum MAXimum DEFault Sets the thermopile accelerator state Queries the thermopile accelerator state Sets the th...

Страница 32: ...gure for sensor presence resistance measurement TEMPerature Configure for sensor temperature measurement MEASure SCALar POWer Performs a power measurement CURRent DC Performs a current measurement VOL...

Страница 33: ...r resist ant Attention To avoid damage to the instrument do not expose it to spray liquids or solvents The unit does not need regular maintenance by the user It does not contain any modules and or com...

Страница 34: ...dBm W cm A Measurement Ranges 106 50 nA 5 mA Ranges selectable in W or A Sensor Dependent Display Resolution 1 pA responsivity value A W AD Converter 16 bit Measurement Uncertainty 0 2 f s 5 A 5mA 0...

Страница 35: ...r depending nm V J Beam Diameter Setting 1 e Sensor Temperature Control Supported Temperature Sensor Thermistor Temperature Measurement Range 10 C 80 C Interface Type USB 2 0 Connector Mini USB left s...

Страница 36: ...nA 50 A 10 nA 5 A 1 nA 500 nA 100 pA 50 nA 10 pA 0 5 f s Voltage Input Thermal Sensors Voltage Range Display Resolution Measurement Accuracy 1 V 100 V 0 5 f s 100 mV 10 V 10 mV 1 V 1 mV 0 1 V 1 0 f s...

Страница 37: ...connection 3 AGND analog ground photodiode ground anode thermal and pyro sensor ground 4 photodiode cathode 5 pyroelectric sensor 8 thermal sensor 7 PRESENT Connect this pin via a 1k 10k resistor to...

Страница 38: ...minimum detectable change of power energy it is not the minimum measurable power energy As the power meter measures a current or a voltage the display resolution can be stated only in A or V In the sp...

Страница 39: ...alibrated and thus able to measure with the specified measurement uncertainty Resolution is the minimum detectable change of the measured parameter The resolution is always specified for a certain pow...

Страница 40: ...the average power of the pulse train is considered Please note that this specification is higher than the max measurable power consequently the sensor will enter saturation and the measurement result...

Страница 41: ...tP pulse duration TP pulse period PPeak pulse peak power Pav pulse average power Repetition rate is related to pulse period Duty cycle is the ratio of pulse duration to pulse period Pulse average powe...

Страница 42: ...he beam shape is not circular the area must be calculated using the appropriate formula For example if the beam shape is elliptical with d1 long axis and d2 short axis the beam area is Average Power D...

Страница 43: ...er W pulse energy tP pulse duration dB beam diameter Pav average power fP repetition rate tP pulse duration dB beam diameter Pulse Energy Density x Given parameters Formula W pulse energy dB beam diam...

Страница 44: ...nd meets all requirements of the Canadian Interference Causing Equipment Standard ICES 003 for digital apparatus These limits are designed to provide reasonable protection against harmful interference...

Страница 45: ...2020 Thorlabs 9 Appendix 43 9 5 Certifications and Compliances...

Страница 46: ...or the technical data of the de scribed unit at any time 9 7 Copyright and Exclusion of Liability Thorlabs has taken every possible care in preparing this document We however assume no li ability for...

Страница 47: ...thorlabs com Thorlabs End of Life Policy WEEE Thorlabs verifies our compliance with the WEEE Waste Electrical and Electronic Equip ment directive of the European Community and the corresponding nation...

Страница 48: ...www thorlabs com...

Отзывы: