Chapter 5: Maintenance 

Page 6 

 

13129-D02

 

Chapter 5  Maintenance 

There  are  no  serviceable  parts  in  the  PDA10DT  detector  or  power  supply.  The 
housing  may  be  cleaned  by  wiping  with  a  soft  damp  cloth.  The  window  of  the 
detector should only be cleaned using isopropyl alcohol and optical grade wipes. 
If you suspect a problem with your PDA10DT, please contact your local Thorlabs 
technical support office and a member of our support team will be happy to assist 
you. 

Chapter 6  Troubleshooting 

Problem 

Suggested Solutions 

There is no signal response. 

Verify that the power is switched on and all 
connections are secure. 

Verify the proper terminating resistor is installed if 
using a Voltage measurement device. 

Verify that the optical signal wavelength is within 
the specified wavelength range. 

Verify that the optical signal is illuminating the 
detector active area 

Ensure, not to block, limit airflow to or stop the 
cooling fan. It will overheat the devise and it will 
affect its functionality. 

Output Voltage will not increase. 
Detector Output is skewed. 

Check to make sure the detector is not saturated. 
Refer to the Output Voltage spec. in the 
Specifications table. 
 
Install a 1" Lens Tube (SM1L10) to the thread 
coupler (SM1T1) to baffle any external light 
sources to see if this improves the response. 

 

 

Содержание PDA10DT

Страница 1: ...PDA10DT EC Amplified InGaAs Detector User Guide...

Страница 2: ...n 4 4 1 Output 4 4 2 Gain Adjustment 4 4 3 Bandwidth Filter Adjustment 5 4 4 Thermoelectric Cooler 5 4 5 Light to Voltage Conversion 5 Chapter 5 Maintenance 6 Chapter 6 Troubleshooting 6 Chapter 7 Spe...

Страница 3: ...cription Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or chassis Terminal Equipotentiality On Supply Off Supply In P...

Страница 4: ...ains It also reduces thermally generated noise The housing acts as a heat sink and includes a fan to increase the cooling capacity It is important to note that the cooling fan will keep the heat sink...

Страница 5: ...DA10DT 4 Plug the power supply into a 47 to 63 Hz 100 to 240 VAC outlet 5 Attach a 50 BNC cable to the output of the PDA When running cable lengths longer than 12 we recommend terminating the opposite...

Страница 6: ...PDA housing When using fiber adapters make sure that the fiber ferrule does not crash into the detector Failure to do so may cause damage to the diode and or the fiber Installing an SM1RR Retaining Ri...

Страница 7: ...loads A 50 series resistor is included on the output to impedance match a 50 coax cable For best performance we recommend operating the PDA10DT with a 50 terminating load located at the end of the coa...

Страница 8: ...temperature drop from the heat sink to the detector element resulting in larger TEC currents Without the fan the TEC current will operate at its limit 820 mA and the detector element will no longer be...

Страница 9: ...al response Verify that the power is switched on and all connections are secure Verify the proper terminating resistor is installed if using a Voltage measurement device Verify that the optical signal...

Страница 10: ...put Impedance 50 Max Output Current 100 mA Load Impedance 50 Hi Z Offset3 20 mV Typ 45 mV Max Offset Drift 70 dB 2 7 mV C TEC Temperature 10 C Bias Voltage 2 V 0 dB and 10 dB 0 V 20 dB 70 dB 1 All mea...

Страница 11: ...dB steps Filter Steps 8 Gain Adjustable Range 70 dB Gain Hi Z 4 Low Pass Filter Bandwidth NEP pW Hz 5 0 dB 1 51 kV A 1M 1 MHz 0 dB 15 9 DC 2 MHz 10 dB 4 75 kV A 500k 500 kHz 10 dB 8 27 DC 1 5 MHz 20...

Страница 12: ...Amplified InGaAs Detector Chapter 7 Specifications Rev M September 7 2021 Page 9 Figure 3 PDA10DT Gain Bandwidth Figure 4 PDA10DT Filter Bandwidth...

Страница 13: ...Amplified InGaAs Detector Chapter 7 Specifications Page 10 13129 D02 Figure 5 Noise at 70 dB Gain amd 500 Hz Bandwidth...

Страница 14: ...Amplified InGaAs Detector Chapter 7 Specifications Rev M September 7 2021 Page 11 Figure 6 Noise at 0 and 70 dB Gain and 1 MHz Bandwidth...

Страница 15: ...Amplified InGaAs Detector Chapter 7 Specifications Page 12 13129 D02 Figure 7 TEC Current vs Temperature Figure 8 Photosensitivity at 10 C and 25 C...

Страница 16: ...Amplified InGaAs Detector Chapter 8 Drawings Rev M September 7 2021 Page 13 Chapter 8 Drawings...

Страница 17: ...rlabs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassembled by t...

Страница 18: ...Amplified InGaAs Detector Chapter 10 Certificate of Conformance Rev M September 7 2021 Page 15 Chapter 10 Certificate of Conformance...

Страница 19: ...USA Canada and South America Thorlabs Inc sales thorlabs com techsupport thorlabs com UK and Ireland Thorlabs Ltd sales uk thorlabs com techsupport uk thorlabs com Europe Thorlabs GmbH europe thorlabs...

Страница 20: ...www thorlabs com...

Отзывы: