THORLABS PDA10DT Скачать руководство пользователя страница 10

Amplified InGaAs Detector

 

Chapter 7: Specifications 

Rev M, September 7, 2021 

Page 7 

Chapter 7  Specifications 

 

 

Specifications

1

 

Optical Specifications 

Wavelength Range 

0.9 - 2.57 µm 

Peak Wavelength (

p

2.3 µm 

Peak Response (

p

1.3 A/W (Typ.)  

Electrical Specifications 

Gain Adjustment Range 

70 dB 

Gain Steps 

Gain Settings (dB) 

0, 10, 20, 30, 40, 50, 60, 70 

Filter Steps 

Filter Settings (kHz) 

0.5, 1, 5, 10, 50, 100, 500, 1000 

Output Voltage

2

 

0 - 5 V 

(50 Ω) 

0 - 10 V (Hi-Z) 

Output Impedance 

50 

Ω 

Max Output Current 

100 mA 

Load Impedance 

50 

Ω - Hi-Z 

Offset

3

  

20 mV (Typ.) 
45 mV (Max) 

Offset Drift (70 dB) 

2.7 mV/°C 

TEC Temperature 

-10 °C 

Bias Voltage 

-2 V (0 dB and 10 dB) 

0 V (20 dB - 70 dB) 

 

 

1

 

All measurements performed with a 50 

Ω load unless stated otherwise. The PDA10DT has a 50 Ω 

series terminator resistor (i.e., in series with amplifier output). This forms a voltage divider with any load 
impedance (e.g., 50 

Ω load divides signal in half). 

2

  Saturation of the output voltage may cause damage to the InGaAs detector element. 

3

  After  the  temperature  has  stabilized  on  all  gain  steps.  Also  note  that  the  worst  case  offset  is  on  the 

10 dB gain step.

 

Содержание PDA10DT

Страница 1: ...PDA10DT EC Amplified InGaAs Detector User Guide...

Страница 2: ...n 4 4 1 Output 4 4 2 Gain Adjustment 4 4 3 Bandwidth Filter Adjustment 5 4 4 Thermoelectric Cooler 5 4 5 Light to Voltage Conversion 5 Chapter 5 Maintenance 6 Chapter 6 Troubleshooting 6 Chapter 7 Spe...

Страница 3: ...cription Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or chassis Terminal Equipotentiality On Supply Off Supply In P...

Страница 4: ...ains It also reduces thermally generated noise The housing acts as a heat sink and includes a fan to increase the cooling capacity It is important to note that the cooling fan will keep the heat sink...

Страница 5: ...DA10DT 4 Plug the power supply into a 47 to 63 Hz 100 to 240 VAC outlet 5 Attach a 50 BNC cable to the output of the PDA When running cable lengths longer than 12 we recommend terminating the opposite...

Страница 6: ...PDA housing When using fiber adapters make sure that the fiber ferrule does not crash into the detector Failure to do so may cause damage to the diode and or the fiber Installing an SM1RR Retaining Ri...

Страница 7: ...loads A 50 series resistor is included on the output to impedance match a 50 coax cable For best performance we recommend operating the PDA10DT with a 50 terminating load located at the end of the coa...

Страница 8: ...temperature drop from the heat sink to the detector element resulting in larger TEC currents Without the fan the TEC current will operate at its limit 820 mA and the detector element will no longer be...

Страница 9: ...al response Verify that the power is switched on and all connections are secure Verify the proper terminating resistor is installed if using a Voltage measurement device Verify that the optical signal...

Страница 10: ...put Impedance 50 Max Output Current 100 mA Load Impedance 50 Hi Z Offset3 20 mV Typ 45 mV Max Offset Drift 70 dB 2 7 mV C TEC Temperature 10 C Bias Voltage 2 V 0 dB and 10 dB 0 V 20 dB 70 dB 1 All mea...

Страница 11: ...dB steps Filter Steps 8 Gain Adjustable Range 70 dB Gain Hi Z 4 Low Pass Filter Bandwidth NEP pW Hz 5 0 dB 1 51 kV A 1M 1 MHz 0 dB 15 9 DC 2 MHz 10 dB 4 75 kV A 500k 500 kHz 10 dB 8 27 DC 1 5 MHz 20...

Страница 12: ...Amplified InGaAs Detector Chapter 7 Specifications Rev M September 7 2021 Page 9 Figure 3 PDA10DT Gain Bandwidth Figure 4 PDA10DT Filter Bandwidth...

Страница 13: ...Amplified InGaAs Detector Chapter 7 Specifications Page 10 13129 D02 Figure 5 Noise at 70 dB Gain amd 500 Hz Bandwidth...

Страница 14: ...Amplified InGaAs Detector Chapter 7 Specifications Rev M September 7 2021 Page 11 Figure 6 Noise at 0 and 70 dB Gain and 1 MHz Bandwidth...

Страница 15: ...Amplified InGaAs Detector Chapter 7 Specifications Page 12 13129 D02 Figure 7 TEC Current vs Temperature Figure 8 Photosensitivity at 10 C and 25 C...

Страница 16: ...Amplified InGaAs Detector Chapter 8 Drawings Rev M September 7 2021 Page 13 Chapter 8 Drawings...

Страница 17: ...rlabs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassembled by t...

Страница 18: ...Amplified InGaAs Detector Chapter 10 Certificate of Conformance Rev M September 7 2021 Page 15 Chapter 10 Certificate of Conformance...

Страница 19: ...USA Canada and South America Thorlabs Inc sales thorlabs com techsupport thorlabs com UK and Ireland Thorlabs Ltd sales uk thorlabs com techsupport uk thorlabs com Europe Thorlabs GmbH europe thorlabs...

Страница 20: ...www thorlabs com...

Отзывы: