![THORLABS MDT693B Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/thorlabs/mdt693b/mdt693b_user-manual_1110769021.webp)
Open-Loop Piezo Controllers
Chapter 7: Remote Communications
Page 20
Rev C, July 7, 2015
The following list shows all available commands and queries, and summarizes
their functions for the MDT694B:
Command
a
Syntax
b
Description
Get Commands
?
List the available commands.
Product Information
id?
Returns the product header and firmware
version
Restore Default
Settings
restore
Restores default factory settings
Get Echo Command
Value
echo?
Returns echo status
Set Echo Command
echo=n
Sets Echo mode. When the echo mode is
on the command parameters will be
returned, or echoed back, to the
communication program.
(Values: 1 = enable; 0 = disable)
Limit Switch Setting
vlimit?
Returns output voltage limit setting.
(75V = 0; 100V = 1; 150V = 2)
Get Display Intensity
intensity?
Returns the display intensity
Set Display Intensity
intensity=n
Set the Display Intensity (0-15).
Read X-Axis Voltage
xvoltage?
Returns the X axis output voltage.
Set X-Axis Voltage
xvoltage=n
Set the output voltage for the X axis.
Read Min. X-Axis
Voltage
Xmin?
Returns the minimum output voltage
limit for X axis.
Set Min. X-Axis Voltage
xmin=n
Sets the minimum output voltage
limit for X axis.
Read System Min.
Voltage
sysmin?
Returns the minimum output voltage
for the system.
Set System Min. Voltage
sysmin=n
Sets the minimum output voltage
for the system.
Read Max. X-Axis
Voltage
Xmax?
Returns the maximum output voltage limit
for X axis.
Set Max. X-Axis Voltage
xmax=n
Sets the maximum output
voltage limit for X axis.
Read System Max.
Voltage
sysmax?
Returns the maximum output
voltage for the system.
Set System Max.
Voltage
sysmax=n
Sets the maximum output voltage
for the system.
Get voltage adjustment
resolution
dacstep?
Reads the current step resolution
Set voltage adjustment
resolution
dacstep=n
Sets the step resolution when
using up/down arrow keys
(n = 1 to 1000)
Increment Voltage
Up arrow
Increments the current channel voltage
by resolution set in dacsteps
Decrement Voltage
Down arrow
Decrements the current channel voltage
by resolution set in dacsteps
Get Friendly Name
friendly?
Returns the friendly name.
Set Friendly Name
friendly=string
Sets the friendly name.
Содержание MDT693B
Страница 1: ...MDT693B and MDT694B Open Loop Piezo Controllers User Guide ...
Страница 30: ...Open Loop Piezo Controllers TTN011225 D02 Rev C July 7 2015 ...