Open-Loop Piezo Controllers
TTN011225-D02, Rev C
July 7, 2015
Master Scan Trim
The
SCAN TRIM
adjusters scale the signals provided by the
MASTER SCAN INT
and
MASTER SCAN EXT
knobs over a range from 80% to 120% on a per-channel
basis. This helps compensate for differences in reactance from one piezo to
another. The
SCAN TRIM
adjusters are only active when Master Scan mode is
active (i.e., the enable LED must be lit).
Note:
The 100% mark is halfway through the adjustment range of the SCAN
TRIM potentiometer. To approximately set the SCAN TRIM to 100%, zero
the pot by turning it counterclockwise as far as possible (a maximum of
four turns). Then adjust the pot clockwise by two turns.
To precisely set the SCAN TRIM to 100%, follow this procedure:
Adjust all of the
INT
control knobs counterclockwise as far as possible.
Disconnect
any
EXT
input connections.
Adjust
all
SCAN TRIM
pots counterclockwise by a minimum of four turns.
Apply a 5 V DC voltage source to the
MASTER SCAN EXT
input.
Adjust
the
SCAN TRIM
pots clockwise until their displays read 75 V
(assuming the voltage limit is set to 150 V).