Doc 2136-D01 Rev E 1/24/02
2-10
2-10
•
Connect the modulation source to the jack "MOD IN" (R1). Avoid ground loops when connecting the function
generator.
If the laser current "ILD" reaches the set current limit "ILIM" in operation, the LED "LIM" (7) lights up, a beeper
sounds a short warning signal, and the laser current is limited to the value of the current limit "ILIM".
If the laser current "ILD" is limited by the selected current limit "ILIM", ripple and noise no longer correspond to the
specifications for normal operation (refer to chapter 2.3.1). However, the set maximum current "ILIM" cannot be
exceeded.
2.3.8
Control Output (Laser Current Monitor)
A rear panle BNC labelled CTL OUT allows the laser current to be monitored with a voltmeter, strip chart recorder,