Chapter 3 Installation
3.2 Mechanical Installation
3.2.1 Environmental Conditions
Location
Indoor use only
Maximum altitude
2000 m
Temperature range
5
o
C to 40
o
C
Maximum Humidity
Less than 80% RH (non-condensing) at 31°C
To ensure reliable operation the unit should not be exposed to corrosive agents or
excessive moisture, heat or dust.
If the unit has been stored at a low temperature or in an environment of high humidity,
it must be allowed to reach ambient conditions before being powered up.
3.2.2 Mounting Options
The K-Cube Piezo Motor Controller is shipped with a baseplate, for use when fitting
the unit to a breadboard, optical table or similar surface.
Warning
The safety of any system incorporating this equipment is the responsibility of
the person performing the installation.
Warning
Operation outside the following environmental limits may adversely affect
operator safety.
Caution
When siting the unit, it should be positioned so as not to impede the operation
of the control panel buttons.
Ensure that proper airflow is maintained to the rear of the unit.