© 2018 Thorlabs
6 Appendix
45
6.2.3 Device Setup Commands
Command List
Command
Description
:WAVEL:VAL
Set the wavelength
:WAVEL:VAL?
Query the currently used wavelength
:WAVEL:RNG?
Query the wavelength range
Description
Set Wavelength
Command syntax:
:WAVEL:VAL <DECIMAL NUMERIC PROGRAM DATA>
Description:
Sets the wavelength [nm] to use for calculating the sensor sensit-
ivity.
Query Wavelength
Command syntax:
:WAVEL:VAL?
Response syntax:
currently used wavelength
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries the currently used wavelength [nm].
Query Wavelength Range
Command syntax:
:WAVEL:RNG?
Response syntax:
minimum settable wavelength [nm]:
<NR3 NUMERIC RESPONSE DATA>,
maximum settable wavelength [nm]:
<NR3 NUMERIC RESPONSE DATA>,
currently used wavelength [nm]:
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries wavelength range.
Содержание ERM100
Страница 1: ...Extinction Ratio Meter ERM100 Operation Manual 2018 ...
Страница 2: ...Copyright 2018 Thorlabs Version Date Item No 1 2 10 Jul 2018 M0009 510 309 ...
Страница 20: ... 2018 Thorlabs 18 ERM100 ...
Страница 22: ... 2018 Thorlabs 20 ERM100 4 1 2 Install Remote Control Application Click I accept if you do so then click Next ...
Страница 24: ... 2018 Thorlabs 22 ERM100 4 1 3 Install Instrument Communicator Click I accept if you do so then click Next ...
Страница 26: ... 2018 Thorlabs 24 ERM100 Click I accept if you do so then click Next ...
Страница 27: ... 2018 Thorlabs 4 Remote Operation 25 Click Next to finish the installation ...
Страница 30: ... 2018 Thorlabs 28 ERM100 Check the box No not this time then Next to continue ...
Страница 33: ... 2018 Thorlabs 4 Remote Operation 31 The logged data can be saved in csv file format ...
Страница 36: ... 2018 Thorlabs 34 ERM100 Select the ERM100 firmware file click Next ...