© 2018 Thorlabs
44
ERM100
6.2.2 Measurement Commands
Command List
Command
Description
:MEAS:VAL?
Query the measurement data
:MEAS:EXT?
Query extreme value monitor
:MEAS:EXR?
Query extreme value monitor and reset extreme values
Description
Measurement Query
Command syntax:
:MEAS:VAL?
Response syntax:
Extinction ratio [dB]
<NR3 NUMERIC RESPONSE DATA>,
Key angle [°]
<NR3 NUMERIC RESPONSE DATA>,
Optical power [dBm]
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries the measurement result.
Note: If a power overflow occurs, the value
‘+Inf’
or
‘-Inf’
is
returned for the optical power value.
Extreme Value Query
Command syntax:
:MEAS:EXT?
Response syntax:
Extinction ratio minimum value [dB]
<NR3 NUMERIC RESPONSE
DATA>,
Key angle maximum value [°]
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries the extreme values of the measurement results. If there
was no new measurement since last monitor reset the values '+Inf'
for the ER and '0.0' for the angle are returned and an error is in-
sert into the error queue.
Extreme Value Query With Reset
Command syntax:
:MEAS:EXR?
Response syntax:
Extinction ratio minimum value [dB]
<NR3 NUMERIC RESPONSE DATA>,
Key angle maximum value [°]
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries the extreme values of the measurement results and re-
sets the extreme value monitor. If there was no new measurement
since last monitor reset the values '+Inf' for the ER and '0.0' for the
angle are returned and an error is insert into the error queue. The
reset sets the ER value to '+Inf' and the angle to '0.0'.
Содержание ERM100
Страница 1: ...Extinction Ratio Meter ERM100 Operation Manual 2018 ...
Страница 2: ...Copyright 2018 Thorlabs Version Date Item No 1 2 10 Jul 2018 M0009 510 309 ...
Страница 20: ... 2018 Thorlabs 18 ERM100 ...
Страница 22: ... 2018 Thorlabs 20 ERM100 4 1 2 Install Remote Control Application Click I accept if you do so then click Next ...
Страница 24: ... 2018 Thorlabs 22 ERM100 4 1 3 Install Instrument Communicator Click I accept if you do so then click Next ...
Страница 26: ... 2018 Thorlabs 24 ERM100 Click I accept if you do so then click Next ...
Страница 27: ... 2018 Thorlabs 4 Remote Operation 25 Click Next to finish the installation ...
Страница 30: ... 2018 Thorlabs 28 ERM100 Check the box No not this time then Next to continue ...
Страница 33: ... 2018 Thorlabs 4 Remote Operation 31 The logged data can be saved in csv file format ...
Страница 36: ... 2018 Thorlabs 34 ERM100 Select the ERM100 firmware file click Next ...