![THORLABS DRV225 Скачать руководство пользователя страница 8](http://html.mh-extra.com/html/thorlabs/drv225/drv225_user-manual_1110741008.webp)
Page 6
ETN043212-D02
DRV225 and DRV250 Stepper Motor Actuators
5) Click the arrow to the right of the window, and select the actuator type (i.e. DRV225 or DRV250) from the list displayed.
Fig. 3.3 Select Actuator Stage Settings Window - 2
6) Click OK.
7) The server reads in the stage and controller information automatically and loads default parameters set at the factory, i.e.
Max speed 20 mm/sec
Max acceleration 20 mm/sec/sec
Home switch Forward.
Refer to the handbook for the associated controller for more information on driving the actuator.
3.4 Maintenance
After prolonged use, and particularly in applications where small movements are continually repeated, the grease on the drive
shaft may build up. This may cause rough or noisy movement, vibration and excessive heating.
It is good practice to run the motor periodically from one end of travel to the other several times in order to redistribute the
grease.
Note
No customer serviceable parts. Return to manufacturer for service.