2-16
A
URORA
C
ONTROL
U
NIT
CAUTION
Keep the PMT Voltage on the Aurora Control Unit set to 0, unless the PMT
is shielded from all sources of light (except the laser during NSOM imaging). Exposure to
light can damage the PMT. The voltage should only be turned up when carefully following
the procedures in Chapter 4.
Figure 2-18
Aurora Control Unit
Power is applied to the Aurora Control Unit (ACU) when the ECU-Plus is powered up.
Failure to observe all CAUTION notices regarding the controls on the ACU may result in
damage to the PMT and probe tip.
The light controls refer to the light on the Aurora-2 head that illuminates the sample area.
This light is only used in procedures that involve the video monitor and should not be on
during NSOM scanning or any time the PMT voltage is turned up.
Note that the Feedback switch is not connected in the current configuration.
PMT
VOLTAGE
PMT
OUTPUT
FEEDBACK
LIGHT
NSOM
AFM
PMT VOLTAGE
CONTROL
LIGHT
BRIGHTNESS
ON
OFF
AURORA CONTROL UNIT