background image

Section 4.2: V1 Mixing Valve Adjustment Instructions with the WWM 

 

Diaphragm Removed

Before beginning repairs read and understand Section 3.1 and follow the steps in Section 3.2 

to decommission the Therm-O-Mix Station. The Station should be tested after all repairs are 

completed.

1. 

Label each hose connection going to the diaphragm housing for easy re-installation. (See 

Appendix #3) 

2. 

Disconnect hoses from the diaphragm housing only. 

3. 

Use a 1-1/2” wrench or spud wrench to loosen diaphragm valve assembly. Once unthreaded, 

pull diaphragm valve out. Some resistance may be felt due to O-ring seal. 

4. 

Remove cap nut and loosen calibration lock nut with a 7/16” wrench. 

5. 

Place the stainless steel sensor into a water bath with a temperature 5F over the set point 

for five minutes. *Note: The sensor should be at or below 75F prior to being placed into the 

water bath for proper calibration. Use calibration screw to adjust the valve spool so that the 

opening between the spool and the cap is .160-.175 in. (See Appendix #5 & #6). 

6.  Place stainless steel sensor into a water bath with a temperature 5°F below the set point for 

five minutes. Verify that the top of the spool is sitting on the bottom of the cap.

7. 

Once properly adjusted, tighten the calibration lock nut and replace cap nut. 

8. 

Thread WWM Diaphragm Actuator back into the V1 Mixing Valve Tee assembly and tighten 

with 1-1/2” wrench. 

9. 

Position the diaphragm housing so that the adaptor fittings are pointing to the side of the 

body with the braided hoses. *Note: To position the diaphragm housing, rotate clockwise 

when looking down on the Therm-O-Mix® Station/WWM to prevent the valve internals from 

loosening. 

10. Reconnect the braided hose lines by hand tightening the hose swivel fittings onto the  

 

 

adaptor fittings. Use a 9/16” open end wrench to tighten the swivel nut 1 to 1-1/2 hex    

 

flats or 60 to 90 past hand tight. *Use second 1/2” wrench as a back-up wrench to

 

 

prevent hose from twisting. 

 

*If braided hose connections containing orifice plates are opened, the orifice plate must be  

 

replaced. Orifice plates have been included in the replacement kit in case they are needed.  

 

(See Appendix #3)

Содержание THERM-O-MIX WWM

Страница 1: ...THERM O MIX WWM STATION Troubleshooting Manual www ThermOmegaTech com 1 877 379 8258...

Страница 2: ...rs V1 Mixing Valve Adjustment Instructions with the Unit Running V1 Mixing Valve Adjustment Instructions with the WWM Diaphragm Removed WWM Diaphragm Actuator VI Mixing Valve Replacement Instructions...

Страница 3: ...ode pressure sensitive diaphragm overrides and closes hot port during over temperature or loss of hot water supply at V1 Page 2 Section 1 2 Recommended Tools Section 1 3 Check List The following is a...

Страница 4: ...en the Station outlet exceeds approximately 95 F thus closing the hot inlet port on the mixing valve With the hot port closed no hot water can flow to the mixing valve so the Station outlet is cold wa...

Страница 5: ...iod of time may be needed depending on the size and length of supply piping 3 With the system activated close the cold water supply valve to the station 4 See Section 4 0 for individual valve adjustme...

Страница 6: ...for repair Remove the V4 Control Valve for testing See section 4 4 If the V4 tests correctly Follow Section 4 2 to adjust the V1 Mixing Valve outside the unit If the valve cannot be adjusted correctl...

Страница 7: ...ol and the cap is 160 175 in See Appendix 5 6 6 Place stainless steel sensor into a water bath with a temperature 5 F below the set point for five minutes Verify that the top of the spool is sitting o...

Страница 8: ...y and tighten with 1 1 2 wrench 7 Position the diaphragm housing so that the adaptor fittings are pointing to the side of the body with the braided hoses Note To position the diaphragm housing rotate...

Страница 9: ...tside connections on the V4 Control Valve require orifice plates to be installed Place an orifice plate in the female end of the hose and carefully thread the hose on to the proper male fitting Note t...

Страница 10: ...or referenced part Appendix 4 The Station checklist should be used whenever troubleshooting or repair work is done on a unit The Station Serial Number is located on the V1 Mixing Body The checklist c...

Страница 11: ...AGM HEAD V1 VALVE PORT C COLD WATER INLET 1 NPT FEMALE V4 VALVE COLD WATER TEPID WATER HOT WATER DIAPHRAGM SUPPLY WATER DIAPHRAGM DRAIN WATER THERMAL SENSING LINE DIAPHRAGM SUPPLY HOSE B DIAPHRAGM SUP...

Страница 12: ...OT WATER AVAILIBLE IS MIXING VALVE OUTLET ABOVE 95 F V4 OPEN DIAPHRAGM ACT OVERRIDE MIXING VALVE V1 BLENDS HOT COLD WATER TO SET POINT IS COLD WATER SUPPLY ABOVE 95 F V4 OPEN DIAPHRAGM ACT OVERRIDE DI...

Страница 13: ...5 ORIFICE ORIFICE ORIFICE V4 CONTROL VALVE V1 MIXING VALVE CAP NUT DIAPHRAGM HEAD PORT B HOT WATER INLET 1 NPT FEMALE FEMALE 1 1 4 NPT HOSE ASSEMBLY 7 5 PORT C COLD WATER INLET 1 NPT FEMALE PORT A TEP...

Страница 14: ...tation supplying tempered water to anything other than the Safety Shower Station YES NO If YES explain __________________________________________________________________________ ______________________...

Страница 15: ...ge 14 APPENDIX 5 WWM Diaphragm Actuator V1 Mixing Valve Line Drawing SET POINT 5 F CARTRIDGE CAP O RING SPOOL THERMOSTATIC ACTUATOR CAP NUT FOR ADJUSTMNET SET SCREW 1 HEX 1 1 2 HEX SET POINT 5 F 160 1...

Страница 16: ...Page 15 APPENDIX 6 WWM Diaphragm Actuator V1 Mixing Valve Adjustment Photo Loosen lock nut Adjust calibration screw...

Страница 17: ...ide WWM Diaphragm Actuator WARNING This product can expose you to chemicals for example lead nickel acrylonitrile which are known to the State of CA to cause cancer birth defects or reproductive harm...

Отзывы: