![Thermoline Scientific STAR-700 Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/thermoline-scientific/star-700/star-700_operating-manuallines_1105751005.webp)
Thermoline Scientific Equipment Pty. Ltd.
T/A
Thermoline Scientific
ABN 80 000 859 129
10-12 Ross Place, Wetherill Park N.S.W. 2164. Australia.
P.O. Box 6862, Wetherill Park Delivery Centre, N.S.W. 1851. Australia.
Phone:
(02) 9604 3911.
International:
61 2 9604 3911.
Fax:
(02) 9725 1706.
International:
61 2 9725 1706.
Email:
Web:
www.thermoline.com.au
Program Set Up Screen Example (Segment 1-5).
The above screen shows an example of a simple five segment program cycle. When starting the
Program, assuming the actual temperature and Humidity is at 25.0degC and 65%RH (the SOV), the
cabinet will take two hours to reach its first target of 10.0degC and 60%RH. The second segment will
hold the 10.0degC and 60% for 10hours. The third segment will repeat segment 1-2 three times. This
will basically hold the 10.0degC and 60%RH for 46hours. The Temperature, Humidity, Co2, Hours,
Minutes, and Seconds Values for Segment 3 are ignored as the segment Step Type is set to ‘Jump
To’. Once it repeats Segment 1-2 three times it will ramp from 10.0degC and 60%RH to 25.0degC and
60%RH over 10 Hours. The Program will then end and control the Safe Operating Values (SOV).
Program Set Up Screen (Segment 1-5). Showing Segment Types