6  |   Operation

SpeedVac™ SPD120

*When the temperature is set to “no”, the heat time is 
defaulted to “0.00” while executing the run.

Saving a program is achieved by pressing and holding down 
the corresponding 

Program

 button for 3 seconds. The 

parameters displayed in the screen are stored into the 
program and can be loaded for use at a later time.

3.

Pre-Heat

 - Use to pre-heat the chamber to 45°C prior to 

beginning a run or between runs. Once a run is initiated, 
the pre-heat automatically stops.

Note that the pre-heat function is only enabled if the 
temperature set point is different than “no”.

4.

Temperature display

 - Indicates the set temperature or 

the actual temperature during a run in °C. The 
temperature set point can be set from 35°C to 80°C at a 
5°C interval. In addition, it is possible to set the 
temperature to “no”, which means that heating will not be 
applied during the run. The temperature setting of “no” 
can be reached under 35°C or over 80°C.

5.

Time display

 - Indicates the heat timer or the run timer 

set points. During a run, it indicates the elapsed or 
remaining run time or the remaining heat time. The range 
for these timers is from 0.01 [1 min.] to 9.59 [9 hours, 59 
minutes] at a 1-minute interval. In addition, it is possible 
to set the heat time to “CCC”, which represents a 
continuous heating throughout the execution of a run. 
The setting of “CCC” can be reached under 0.01 [1 min.] 
or over 9.59 [9 hours, 59 minutes].

6.

Select

 - Press this button to select the parameters to be 

modified. The selection is from left to right, in a cyclic 

manner. The selected parameters is shown by the 

corresponding indicator that is illuminated.

7.

Minus (“-”)

 - Decrements the value of the selected 

parameter.

8.

Plus (“+”)

 - Increments the value of the selected 

parameter.

9.

View

 - During a run, pressing view shows the set run 

parameters. When no run is ongoing, pressing view 
displays the actual live parameters. The time display 
shows either “0.00” or the end status of the last executed 
run (e.g. “End”, “Err.”).

10.

Manual run

 - Starts a “Manual” run based on the 

parameters loaded in the screen, use manual run 
function if you do not want to set a fixed run time.

11.

Auto run

 - Starts an “Automated” run based on the 

parameters loaded in the screen, use auto run function if 
you want to set a fixed run time.

12.

Stop

 - Terminates a “Manual” or 

Auto

 run.

Note: 

Restoring the factory default parameters for a given 

program is achieved by selecting it and then pressing and 
holding down the 

Stop

 button for 3 seconds. The default 

parameters associated with the program (factory default) will 
be restored and the screen updated accordingly.

Audible Notification

The concentrator will play tones depending on the user 
actions. Below is a description of these tones. 

Table 2. Program Settings

Program

Description

Default UI 
parameters

Program 1

Water

Temperature: 45°C

Heat time: 2.00

Run time: 2.00

Program 2

“Low Boiling”   
ACN/MEOH

Temperature: 65°C

Heat time: 2.00

Run time: 2.00

Program 3

HPLC

H2O + Low Boil 
s.1% acid

Temperature: “no”

Heat time: 0.01*

Run time: 2.00

Note: 

The programs are for convenience only. 

Prior to use, users are advised to conduct a risk 
assessment to confirm that the program 
parameters meet the user’s application specific 
needs and conditions. Users accept the risks 
associated with using these programs.

Table 3. Audible Notifications

Event

Sound Description

Start-up

5 beeps

Starting a run (manual 
or automatic)

1 single beep

Manually stopping a 
manual or automatic 
run

3 beeps

Automatically 
stopping an automatic 
run

1 beep that is repeated until the 
user pushes the 

Stop

 button or 

opens the cover of the 
concentrator

Program saved

3 beeps

Содержание ThermoScientific SpeedVac SPD120

Страница 1: ...Installation and Operation 80302132 Revision A July 2018 SpeedVac SPD120 Vacuum Concentrator ...

Страница 2: ...rmo Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited except with the written authorization of Thermo Fisher Scientific Inc The contents of this document are subject to change without notice All technical info...

Страница 3: ...ions 3 Installation 4 Contents 4 Installation of SPD120 4 Operation 5 Description Of Control Panel 5 Manual Run 7 Auto Run 7 Drying Rate 8 Exporting Live Run Data 8 HyperTerminal Configuration 8 Lid Stay Mechanism 9 Other Topics 9 Rotors 10 Accessories 11 Appendix 12 Additional SPD120 System Setup 12 Warranty 13 ...

Страница 4: ...tion trap available separately for virtually 100 solvent recovery This results in a thorough evaporation process in a safe environment for both the sample and the user The unit features an advanced front panel with dual timer for automatic or manual control of run conditions it also features three programs that are modifiable which enable user to retrieve and store application protocols for future...

Страница 5: ...dicates extreme low temperatures and high risk of frostbite Do not touch bare metal or samples with unprotected body parts This symbol indicates possible pinch points which may cause personal injury This symbol indicates a need to use gloves during the indicated procedures If performing decontamination procedures use chemically resistant gloves Before installing using or maintaining this product p...

Страница 6: ... per Installation Categories II Operative Power 115 VAC 60 Hz 6 amps 230 VAC 50 Hz 3 amps Vacuum Chamber TEFLON coated chemical resistant aluminum casting Cover Glass cover includes a safety interlock Induction Motor Maintenance free Bearings Standard Chamber temperature 35 C to 80 C 5 C increments Weight 62 lbs 28 kg Dimensions W x D x H 14 in x 18 in x 14 in 36 cm x 46 cm x 36 cm Fuse 6 A 250 VA...

Страница 7: ...6 amps The SPD120 units configured for 230 VAC 50 Hz should be plugged into a circuit rated for at least 3 amps Note Do not use any detachable power cord that is not adequately rated for the unit Installation of SPD120 Figure 1 Rear View of System Table 1 Contents Qty Description 1 Thermo Scientific Savant SpeedVac SPD120 Concentrator 1 Line Cord 1 To prevent damage to pump due to vapor exposures ...

Страница 8: ... pin with the groove Secure the assembly by screwing the retaining knob into the drive shaft above the rotor Tighten it firmly but not excessively Description Of Control Panel 1 Lid locked Indicates if the lid of the concentrator is locked or not The indicator will be illuminated when locked 2 Program buttons 3 programs that are modifiable Loading a program is achieved by pressing the correspondin...

Страница 9: ... the value of the selected parameter 9 View During a run pressing view shows the set run parameters When no run is ongoing pressing view displays the actual live parameters The time display shows either 0 00 or the end status of the last executed run e g End Err 10 Manual run Starts a Manual run based on the parameters loaded in the screen use manual run function if you do not want to set a fixed ...

Страница 10: ... bleed into the chamber 11 After the rotor stops spinning the cover lock disengages and the lid locked indicator is turned off The display reverts to last set parameters 12 Open the cover and remove the samples GENERAL During the run display shows actual parameters To check set parameters press View button and Select button The display will revert temporarily to the set points for approximately 5 ...

Страница 11: ...he data are sent at a fixed baud rate of 115200 and are comma separated as shown in the format temperature heat time run time The temperature is represented as a 2 digit integer in C The heat time is expressed as a 3 digit integer in minutes or CCC for continuous heating The runtime is expressed as a 3 digit integer in minutes Note The following section details the step by step procedures for conf...

Страница 12: ... of the run when the lid has already been locked the unit will automatically abort the run and signal the user with the Lid message on the display and an error tone Otherwise during a run the user cannot open the cover due to vacuum in the concentration chamber The cover is locked down at all times during a run and whenever power to the unit is interrupted The cover lock is an additional safety fe...

Страница 13: ...ances CAUTION Product Maintenance Maintenance of product only permitted to be performed by trained service staff using suitable measuring and auxiliary equipment as well as detailed service instruction Solvent Solvent Combination Wear gloves eye protection masks and lab coats while working on the system WARNING Burns due to hot device parts When the concentrator is set to a temperature of 60 C and...

Страница 14: ...2MP 6 Shallow well plates RHSW6MP Sample Format Type Working Volumes ml Number Description Traditional Rotor Model Part No Description GCF400 Glass condensation flask FC400 Flask cover for GCF400 145 6012 00 Foam insulating ring SCC1 CryoCool heat transfer fluid 1 liter ANT100 Post trap assembly for Oligo Preps Used on UVS850DDA only ANS121 Ammonia neutralizing solution for Oligo Preps Used on UVS...

Страница 15: ... Figure 6 Component High Vacuum System SpeedVac Concentrator SPD120 Vacuum Tubing To Pump UVS450DDA ANT100 Exhaust recovery Post Trap Bleeder Valve Universal Vacuum Source SpeedVac Concentrator SPD120 Flexible Vacuum Tubing RVT450 Cold Trap SCT120 DVG50 VLP120 Vacuum Pump Chemical Trap Bleeder Valve ...

Страница 16: ...MA number issued by Thermo Costs of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this manual When returning apparatus that may contain hazardous material you must pack and label them following U S Department of Transportation DOT regulations applying to transportation of hazardous materials Your shipping documents must also meet DOT regulat...

Страница 17: ...to tramite queste medesime Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con queste Direttive l elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito www thermofisher com WEEERoHS in Servizi e Supporto Conformité WEEE Les prod...

Страница 18: ... 30 France 33 2 2803 2180 33 2 2803 2180 Germany 0800 1 536 376 49 6184 90 6000 0800 1 536 376 Italy 32 02 95059 552 39 02 95059 552 432 254 375 Netherlands 31 76 579 55 55 31 76 571 4440 Nordic Baltic CIS 358 9 329 10200 358 9 329 100 Russia 7 812 703 4215 7 812 703 4215 Spain Portugal 34 93 223 09 18 34 93 223 09 18 Switzerland 41 44 454 12 22 41 44 454 12 12 UK Ireland 44 870 609 9203 44 870 60...

Отзывы: