ThermoFisher Scientific ThermoScientific SpeedVac SPD120 Скачать руководство пользователя страница 8

SpeedVac™ SPD120

 Operation  |   5

Operation

Connect the power cord to the power inlet on the back of the 
unit and plug it into the appropriate outlet. Turn on the main 
power switch located on the front of the unit in the lower right 
to energize the unit. The safety lid locking mechanism on the 
SPD120 will automatically disengage. Do not attempt to open 
the lid unless the unit is plugged in.

Note: 

The SP120 Concentrator is an important component in 

a complete concentration system that should also include a 
refrigerated vapor trap and a vacuum pump along with other 
applicable accessories. Please refer to operating instructions 
of each component for details on usage.

Rotor installation.

 Open the lid of the rotor chamber. 

Visually align the pin on the drive shaft with the groove on the 
bottom of the rotor. Carefully lower the rotor onto the drive 
shaft. Rotate the rotor by hand to ensure alignment of the pin 
with the groove. Secure the assembly by screwing the 
retaining knob into the drive shaft above the rotor. Tighten it 
firmly but not excessively. 

Description Of Control Panel

1.

Lid locked 

- Indicates if the lid of the concentrator is 

locked or not. The indicator will be illuminated when 
locked.

2.

Program buttons

 - 3 programs that are modifiable.

Loading a program is achieved by pressing the 
corresponding 

Program

 button. The following 

parameters are automatically loaded onto the screen:

Temperature set point

Heat time

Run time

Default Program Settings

The table 2 below shows the factory default values attributed 
to the programs parameters.

CAUTION: 

Use appropriate Thermo Fisher 

Scientific designed rotors recommended for the 
product.

CAUTION: 

Load the rotor and close the lid. 

Always balance rotor loads. An imbalanced rotor 
causes vibration that will damage the system’s 
bearings and equipment. Load the rotor 
symmetrically. There need not be a tube in each 
holder, but the load must be evenly distributed. 
When using a rotor with aluminum tube holders, 
insert all the tube holders. 

1

3

4

5

10

11

12

2

6

7

8

9

Содержание ThermoScientific SpeedVac SPD120

Страница 1: ...Installation and Operation 80302132 Revision A July 2018 SpeedVac SPD120 Vacuum Concentrator ...

Страница 2: ...rmo Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited except with the written authorization of Thermo Fisher Scientific Inc The contents of this document are subject to change without notice All technical info...

Страница 3: ...ions 3 Installation 4 Contents 4 Installation of SPD120 4 Operation 5 Description Of Control Panel 5 Manual Run 7 Auto Run 7 Drying Rate 8 Exporting Live Run Data 8 HyperTerminal Configuration 8 Lid Stay Mechanism 9 Other Topics 9 Rotors 10 Accessories 11 Appendix 12 Additional SPD120 System Setup 12 Warranty 13 ...

Страница 4: ...tion trap available separately for virtually 100 solvent recovery This results in a thorough evaporation process in a safe environment for both the sample and the user The unit features an advanced front panel with dual timer for automatic or manual control of run conditions it also features three programs that are modifiable which enable user to retrieve and store application protocols for future...

Страница 5: ...dicates extreme low temperatures and high risk of frostbite Do not touch bare metal or samples with unprotected body parts This symbol indicates possible pinch points which may cause personal injury This symbol indicates a need to use gloves during the indicated procedures If performing decontamination procedures use chemically resistant gloves Before installing using or maintaining this product p...

Страница 6: ... per Installation Categories II Operative Power 115 VAC 60 Hz 6 amps 230 VAC 50 Hz 3 amps Vacuum Chamber TEFLON coated chemical resistant aluminum casting Cover Glass cover includes a safety interlock Induction Motor Maintenance free Bearings Standard Chamber temperature 35 C to 80 C 5 C increments Weight 62 lbs 28 kg Dimensions W x D x H 14 in x 18 in x 14 in 36 cm x 46 cm x 36 cm Fuse 6 A 250 VA...

Страница 7: ...6 amps The SPD120 units configured for 230 VAC 50 Hz should be plugged into a circuit rated for at least 3 amps Note Do not use any detachable power cord that is not adequately rated for the unit Installation of SPD120 Figure 1 Rear View of System Table 1 Contents Qty Description 1 Thermo Scientific Savant SpeedVac SPD120 Concentrator 1 Line Cord 1 To prevent damage to pump due to vapor exposures ...

Страница 8: ... pin with the groove Secure the assembly by screwing the retaining knob into the drive shaft above the rotor Tighten it firmly but not excessively Description Of Control Panel 1 Lid locked Indicates if the lid of the concentrator is locked or not The indicator will be illuminated when locked 2 Program buttons 3 programs that are modifiable Loading a program is achieved by pressing the correspondin...

Страница 9: ... the value of the selected parameter 9 View During a run pressing view shows the set run parameters When no run is ongoing pressing view displays the actual live parameters The time display shows either 0 00 or the end status of the last executed run e g End Err 10 Manual run Starts a Manual run based on the parameters loaded in the screen use manual run function if you do not want to set a fixed ...

Страница 10: ... bleed into the chamber 11 After the rotor stops spinning the cover lock disengages and the lid locked indicator is turned off The display reverts to last set parameters 12 Open the cover and remove the samples GENERAL During the run display shows actual parameters To check set parameters press View button and Select button The display will revert temporarily to the set points for approximately 5 ...

Страница 11: ...he data are sent at a fixed baud rate of 115200 and are comma separated as shown in the format temperature heat time run time The temperature is represented as a 2 digit integer in C The heat time is expressed as a 3 digit integer in minutes or CCC for continuous heating The runtime is expressed as a 3 digit integer in minutes Note The following section details the step by step procedures for conf...

Страница 12: ... of the run when the lid has already been locked the unit will automatically abort the run and signal the user with the Lid message on the display and an error tone Otherwise during a run the user cannot open the cover due to vacuum in the concentration chamber The cover is locked down at all times during a run and whenever power to the unit is interrupted The cover lock is an additional safety fe...

Страница 13: ...ances CAUTION Product Maintenance Maintenance of product only permitted to be performed by trained service staff using suitable measuring and auxiliary equipment as well as detailed service instruction Solvent Solvent Combination Wear gloves eye protection masks and lab coats while working on the system WARNING Burns due to hot device parts When the concentrator is set to a temperature of 60 C and...

Страница 14: ...2MP 6 Shallow well plates RHSW6MP Sample Format Type Working Volumes ml Number Description Traditional Rotor Model Part No Description GCF400 Glass condensation flask FC400 Flask cover for GCF400 145 6012 00 Foam insulating ring SCC1 CryoCool heat transfer fluid 1 liter ANT100 Post trap assembly for Oligo Preps Used on UVS850DDA only ANS121 Ammonia neutralizing solution for Oligo Preps Used on UVS...

Страница 15: ... Figure 6 Component High Vacuum System SpeedVac Concentrator SPD120 Vacuum Tubing To Pump UVS450DDA ANT100 Exhaust recovery Post Trap Bleeder Valve Universal Vacuum Source SpeedVac Concentrator SPD120 Flexible Vacuum Tubing RVT450 Cold Trap SCT120 DVG50 VLP120 Vacuum Pump Chemical Trap Bleeder Valve ...

Страница 16: ...MA number issued by Thermo Costs of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this manual When returning apparatus that may contain hazardous material you must pack and label them following U S Department of Transportation DOT regulations applying to transportation of hazardous materials Your shipping documents must also meet DOT regulat...

Страница 17: ...to tramite queste medesime Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con queste Direttive l elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito www thermofisher com WEEERoHS in Servizi e Supporto Conformité WEEE Les prod...

Страница 18: ... 30 France 33 2 2803 2180 33 2 2803 2180 Germany 0800 1 536 376 49 6184 90 6000 0800 1 536 376 Italy 32 02 95059 552 39 02 95059 552 432 254 375 Netherlands 31 76 579 55 55 31 76 571 4440 Nordic Baltic CIS 358 9 329 10200 358 9 329 100 Russia 7 812 703 4215 7 812 703 4215 Spain Portugal 34 93 223 09 18 34 93 223 09 18 Switzerland 41 44 454 12 22 41 44 454 12 12 UK Ireland 44 870 609 9203 44 870 60...

Отзывы: