6.4.2 Electrical Power Supply Connection
WARNING!
The ARL EQUINOX is equipped with a standard IEC 60320/C13 standard connector
used also by computers.
For french and US sockets, the power supply cord is shipped with the instrument.
For other socket types, the power supply cord has to be obtained locally.
The socket providing the electric power supply for the instrument must have a proper ground
connection.
6.4.3 Connection Gas
SERVICE!
Open the valve (2) of the gas cylinder (1) and open slightly the gas regulator (3) in
order to have a low gas flow. Then, connect the male quick f
itting to the “Inside Gas
Bottle” connector (4). This will avoid air pollution in the detector.
Adjust very slowly the outlet gas pressure on the
regulator to the recommended value (6 bar). A
gentle flush can be performed on connection (5). If
the gas cylinder cannot be installed as a part of the
instrument, please follow the instructions in Section
Installation of an External Gas Cylinder on page