Chapter |
Connection Specifications
Confidential, limited rights
PN AA83838 | Revision C | 11-Mar-2019
ARL EQUINOX 100
Page
17
ARL EQUINOX X-Ray diffractometer can be equipped with a gas cylinder integrated directly
to the instrument. In this case, the cylinder should not exceed given dimensions based on the
specific instrument. Our gas regulator is equipped with a safety valve adjusted to 6 .5 Bar. In
case of overpressure, a tubing allows to release the gas outside of the instrument.
Note
Always use a gas regulator with the instrument.
Maximum dimension of the gas cylinder:
Diffractometer
model
Height
– Valve
included- (mm)
Diameter (mm)
Recommended Gas Regulator
ARL EQUINOX 100
380
85
Messer (type FM53 with indicators
positioned next to each other)
If the gas cylinder cannot be installed as a part of the instrument, please follow the
instructions in Section Installation of an External Gas Cylinder on page
5.3.1 Specifications of the Detector Gas
Specification Parameter
Value, Comments
Gas
Argon / Ethane (C
2
H
6
)
Argon: Q.S. N56
Ethane : 15%+/-0,3% Ethane N35
Detector high voltage
9.8 kV
Gas Pressure
6 Bar
Gas Flow
450 mm
/min +/- 100 mm
/min (i.e. 2 bubbles/min)