Chapter 3 | Setup Menu
3.2.7 Calibration Buffer Set
Set the pH buffer set as USA or DIN for automatic buffer
recognition during pH calibrations.
USA: 1.68, 4.01, 7.00, 10.01, 12.46
DIN: 1.68, 4.01, 6.86, 9.18, 12.46
Note: During a pH calibration, pH buffer values can also be
manually entered for each calibration point.
Press the
or
key to scroll through the Setup
Menu list until the Calibration Buffer Set item is
shown.
Press the
or
key to set the pH buffer set.
Press the “Enter” key to save the change.
3.2.8 Read Type
Set the Read Type as Continuous, Auto-Read or Timed to
define how measurements are performed and when
measurements are saved and exported.
Continuous:
In the Continuous measure mode, measurements are
continuously updated on the display and the stopwatch
(stabilizing) or checkmark (stable) icon indicates the
measurement stability status. Press the “Log/Export” key
to save a measurement to the data log and export.
Auto-Read:
In the Auto-Read measure mode, press the “Measure” key
to start a measurement. When the measurement is stable,
the checkmark (stable) icon is shown and the
measurement is locked on the display until the “Measure”
key is pressed again. The stable measurement is
automatically saved to the data log and exported.
Timed:
In the Timed measure mode, the measurement values are
continuously updated on the display. Measurements are
automatically saved to the data log and exported at the
selected time intervals, from 5 seconds to 60 minutes, the
entire time the meter is in the measurement mode.
Press the
or
key to scroll through the Setup
Menu list until the Read Type item is shown.
Press the
or
key to set the read type.
Press the “Enter” key to save the change.