Tighten the cover by screwing the 8 x M4 screws.
Tighten in diagonal to homogenize the pressure around
the cover.

Preparing the Sample Inside a Glove Box

All  this  procedure  must  be  performed  inside  an  appropriate  glove  box  if  sample  is  air  sensitive  or
presents any risk. In this case, be careful to clean all metallic parts to avoid any pollution of the glove
box. The cell should be opened.

Do not forget any components:

Main core of the sample holder,
Support of sample holder,
Cover and 8 screws and Allen keys,
Crucible,
O-ring,
Spatula and glass slide,
sample!

Put the sample holder back on the glove box airlock. Install on the ARL diffractometer as described
above.

If sample does not require any further preparation, then proceed as described in 

Preparing the Sam-

ple Outside a Glove Box

 on page 9

 .

If the sample requires a preparation, such as gas flushing or gas replacement, then follow the next
steps.

Preparing Sample Inside a Glove Box and Gas Replacement

If the cell should be connected to gas supply or vacuum pump, then it is necessary to connect the
valve interface between SSCA and your gas exchanger device.

Valve adaptor (SSCA-VA) to connect SSCA to external device (pumping, flushing…)

With this fitting, it is possible to connect the sample holder to a gas cylinder or a vacuum pump when
SSCA is extracted from glove box.

Confidential, limited rights

PN AA83871 | Revision A

SSCA Sample Holder: Instruction Notice

Page

 10

Sample Preparation Procedure

Chapter|

5.2







5.3

Содержание ARL EQUINOX Series

Страница 1: ...Thermo Scientific ARL EQUINOX Series SSCA Sample Holder Instruction Notice AA83871 A February 2022...

Страница 2: ...to the Power Supply 4 3 Setting of the Incidence Angle 5 Sample Preparation Procedure 5 1 Preparing the Sample Outside a Glove Box 5 2 Preparing the Sample Inside a Glove Box 5 3 Preparing Sample Ins...

Страница 3: ...his document is the original manual The information in this document is subject to change without notice We assume no responsibility for any errors that may appear in this document Confidential limite...

Страница 4: ...nal information contact Thermo Fisher Scientific Supplied Items Before the shipment the sample holders are tested and packed carefully However damage can oc cur during the transport If the sample hold...

Страница 5: ...not specified by Thermo Fisher Scientific the protection provided by the sample holder may be impaired Only accessories that comply with Thermo Fisher Scientific specifications are allowed to be used...

Страница 6: ...strapping of a Kapton window on a Viton O ring and by a stainless steel belt This makes the easy replacement of the window Two rapid connectors allow to connect to the gas system like vacuum pump of...

Страница 7: ...er has to be inser ted in the reference hole For more detailed information refer to the ARL EQUINOX instruction notice Connections to the Power Supply The SSCA sample holder is supplied by 12VDC The p...

Страница 8: ...in the picture below The longest line represents the incidence angle equal to 0 2 Each short line represents 2 of incidence angle 4 The medium long lines represent 10 of inci dence angle 3 Then the i...

Страница 9: ...he surface of the powder mount in order to avoid line displacement error due to height error Use a glass slide to make the surface flat and homogeneous Set the Kapton disk in the housing to make ease...

Страница 10: ...ractometer as described above If sample does not require any further preparation then proceed as described in Preparing the Sam ple Outside a Glove Box on page 9 If the sample requires a preparation s...

Страница 11: ...ing for low energy X ray it is better to replace the argon gas by heli um nitrogen or other inert acceptable gas Introduction of components inside the glove box The following figure shows the parts wh...

Страница 12: ...le holder exit from the glove box Insertion of the SSCA in the ARL EQUINOX Insert the sample holder in the sample holder mounting hole of the instrument Do not use any un necessary force The sample ho...

Страница 13: ...ing the sealing between the cover and the sample holder should be replaced if damaged No glue Window Ref 100421014 can be replaced if necessary Remove the 24 M3 screws Remove the Kapton foil Remove an...

Страница 14: ...in ON position If the problem persists contact Thermo Fisher Scientific You observe peak shift Check your sample preparation make sure that you have right amount of powder in the cup Check the alignm...

Страница 15: ...irectives EN 61010 1 2010 Safety requirements for electrical equipment for measurement control and laboratory use Part 1 General requirements 2011 65 EC restriction of the use of certain hazardous sub...

Страница 16: ...ng Sample Inside a Glove Box and Gas Replacement 10 Preparing the Sample Inside a Glove Box 10 Preparing the Sample Outside a Glove Box 9 R Revision Table 3 S SSCA Sample Holder Description 5 Safety 5...

Страница 17: ...ic and its subsidiaries unless otherwise specified Learn more at thermofisher com elemental or email us at info spectrometry thermofisher com Thermo Fisher Scientific Brno s r o Vlastimila Pecha 1282...

Отзывы: