OM6464K Rev. 4    11 January 2004

Page 9 of 24

CAUTION

!

If the rotor chamber is not at the specified 

temperature, it will NOT abort a spin. Press the 
STOP key if you need to stop a run. 
At the beginning of a spin, if the chamber 

temperature differs by more than 5° C from the 
specified temperature, the temperature display will 
switch between the actual and set/programmed 

temperatures until the two temperatures come to 
within 5° C.  

Main Power Switch

The main power switch is located on the left side of the 
centrifuge, next to the power cord.  Green= ON, 

RED=OFF.

Circuit Breaker

The system provides an automatic circuit breaker for 
emergency situations such as power surges that could 
damage the unit.  The breaker is integral to the power 

switch.  When tripped, the red (power off) button will 
protrude from the power switch.

• If the circuit breaker trips, press the green button in the 

power switch to reset the breaker.

3.6 Front Panel

IEC Centra CL5R Front Control Panel
The control panel contains numeric displays for all 
operational controls. These displays have two states or 

modes: Actual (bright display) and Set (dim display).  

Actual Mode

In Actual mode (bright display), the display indicates 
current run conditions, such as:  

• rotor speed or force
• elapsed time or time remaining in the run
• actual temperature (Refrigerated only)

• program number

Set Mode

In Set mode (dim display), the display indicates the desired 
settings for the run.  Set mode is activated when SPEED, 
TIME, TEMP., or PROGRAM are pressed, briefly at the start 

of a run, or briefly after the unit is switched ON.
The numeric displays also display warning or error 

messages

.

1 2 3
4 5 6
7 8 9

CE

0/-

ENTER

RPM

RCF

HOLD

HOLD

TIME

SPEED

RADIUS

ADJUST

START

STOP/LID

ACCEL

DECEL

SHORT

SPIN

TEMP

PROGRAM

SAVE

MANUAL

°C

RPM
RCF

Control/

Indicator

Function

SPEED

Switches the display from Actual to Set mode. Select 
the desired speed using the numeric keypad then press 
ENTER. A selection must begin before 5 seconds elapse 
or the display will revert to the Actual reading. 

RPM

RCF

The number in the speed/force display represents the 
rotor speed in RPM or force in RCF.  Press this key to 
toggle between RPM and RCF.  When RPM is selected 
the display indicates revolutions per minute.  When RCF 
is selected the display indicates relative centrifugal force 
in units of g.  Use the numeric keys to change the set 
speed or force. Select speed in increments of 50 RPM, 
from 500 through 15,000 RPM (depending on the max. 
allowable speed limit for the particular rotor).  Select 
RCF from 500 - 21,000 xg (depending on the max. 
allowable RCF limit for the particular rotor) in increments 
of 50 xg.

ROTOR 

RADIUS

SPEED & RPM/RCF:

  

Press these keys together to 

change the radius of rotation.  Select the radius using 
the numeric keypad then press ENTER.  Also see Section 
3.4 Rotor Recognition System. 

TIME

Switches the display from Actual to Set mode. Select 
the desired time using the numeric keypad then press 
ENTER. A selection must begin before 5 seconds elapse 
or the display will revert to the Actual reading. 

Time display

: The number in the display indicates time. 

Time is displayed as minutes:seconds up to 99 minutes, 
59 seconds.  The timer begins counting at the start of a 
run, for example, when START is pressed.  In manual 
mode the timer counts down from the set point.  In time 
HOLD or SHORT SPIN modes

,

 the timer counts up. 

HOLD

The hold feature is used to initiate an indefinite spin.  
Press the HOLD key so that HOLD appears in the display.  
Pressing START will then begin a run at the set 
parameters. The timer counts up and the run will not end 
until the STOP key is pressed.

SHORT SPIN

The short spin is used for quick separations. Pressing 
SHORT SPIN will begin a run at the set parameters.  The 
timer counts up and the run will not end until the SHORT 
SPIN key is released.

TEMP

(CL5R only)

The TEMP key switches the display from Actual to Set 
mode. Select the desired temperature using the numeric 
keypad, and then press ENTER. A selection must begin 
before 5 seconds elapse or the display will revert to the 
Actual reading.

Temperature display: The number in the display 
represents temperature in degrees Celsius, from 

9

° 

C through 

+

40

° 

C. 

Note:

 If the actual rotor chamber temperature differs by 

more than 5° C from a selected set point temperature, 
the °C display will switch between the actual and set/
programmed temperatures, until the two temperatures 
come to within 5° C. 

PROGRAM 

Switches the display from Manual to Program mode.

SAVE

Saves the currently displayed desired settings as stored 
programs 1 through 99 (see Section 3.3).  The numeric 
display shows the stored program number and mode of 
operation (see Section 3.2).

MANUAL

Returns the unit to manual operation from the Program 
Mode.

Содержание 6464K

Страница 1: ...EC Centra CL5 and CL5R Centrifuges OM6464K Rev 4 11 January 2004 CL5 VENTILATED Cat No 6464K 120 VAC 60Hz Cat No 6465K 220 240 VAC 50 60Hz CL5R REFRIGERATED Cat No 6466K 120 VAC 60Hz Cat No 6467K 220 240 VAC 50 60Hz Instruction Manual ...

Страница 2: ......

Страница 3: ... 1 Rotor and Accessories 6 3 2 Starting and Stopping a Run 7 3 3 Stored Programs 8 3 4 Rotor Recognition System 8 3 5 Refrigeration IEC Centra CL5R only 8 3 6 Front Panel 9 3 7 Diagnostic Messages and Error Codes 10 4 Applications 11 4 1 Introduction 11 4 2 Speed and Force Tables 12 4 3 Derating Tables for Dense Samples 14 4 4 Chemical Resistance Table 15 4 5 Decontamination Table 16 4 6 RCF Nomog...

Страница 4: ...OM6464K Rev 4 11 January 2004 Page 4 of 24 ...

Страница 5: ...nual is written for centrifuge operators In addition to operation information it contains a few basic troubleshooting techniques and a chapter on maintenance This manual is not a guide for servicing centrifuge units Warnings Cautions and Notes WARNING WARNING A WARNING advises against certain actions or situations that could result in personal injury CAUTION A CAUTION advises against actions or si...

Страница 6: ...its new location ensure that the suction cups adhere correctly to the work surface 3 Operation 3 1 Rotor and Accessories A balanced load is essential for all centrifuges An unbalanced load produces vibration and can damage the unit ALWAYS balance containers on opposite sides of the rotor NOTE A 2 gram load imbalance at a speed of 4600 RPM imparts force equivalent to 20 pounds 9 1 kg at rest Always...

Страница 7: ...ions 1 5 and 10 14 or 7 3 and 16 12 3 Load six tubes at positions 1 9 5 and 14 18 10 or 7 9 3 and 12 18 16 4 Loading an odd number of tubes is not recommended Rotor Removal To remove the rotor 1 If your rotor has a dic remove this first by loosening the nut 2 Using the wrench provided loosen the rotor locking nut 3 Remove the wrench 4 Remove the locking nut 5 Lift the rotor off the shaft 3 2 Start...

Страница 8: ...ther with its default radius if not explicitly entered to the program Recall Program 1 Press the PROGRAM key to enter program mode 2 Select the appropriate program number by entering the desired program number on the numeric key pad and press ENTER The program s set parameters will display 3 Press START to begin this run Lock Program Unlock Program To lock a program to prevent changes to the progr...

Страница 9: ...lay indicates relative centrifugal force in units of g Use the numeric keys to change the set speed or force Select speed in increments of 50 RPM from 500 through 15 000 RPM depending on the max allowable speed limit for the particular rotor Select RCF from 500 21 000 xg depending on the max allowable RCF limit for the particular rotor in increments of 50 xg ROTOR RADIUS SPEED RPM RCF Press these ...

Страница 10: ... 3 7 Diagnostic Messages and Error Codes The beeper sounds in these situations Two times on power up Three times at the end of a spin Five times when a warning occurs Diagnostics Diagnostic messages appear in place of the speed display in the following cases Message Description bAL bAL indicates an unbalanced rotor Open the cover to erase this message Verify that a balanced load is installed Inspe...

Страница 11: ... Direction of Rotation The microprocessor discovered wrong direction of rotation during acceleration Warnings during a spin LId PFAIL and dIR error messages can occur during a spin In this case the rotor brakes or coasts to a stop and the run ends 4 Applications 4 1 Introduction This section describes the use of specific rotors and accessories More detailed information is shipped with the rotor or...

Страница 12: ...6 2 pr x 6559E 4 x 100 ml sealed Round bottom 38x111 mm 4500 3940 17 4 2 pr x 6557E 8 x 50 mL sealed Corning 8240 50 Thermo Electron 2997 4500 3980 17 6 2 pr x 6556E 8 x 50 mL sealed Falcon Corning conical 4500 3980 17 6 2 pr x 6560E 12 x 25 30 mL sealed Universal 25 x 107mm 4500 3400 15 0 2 pr x 6568E 20 x 15 mL sealed Falcon Corning conical 4500 3980 17 6 2 pr x 6561E 20 x 10 ml sealed Kova Uris...

Страница 13: ...earest 50 rpm or 50 xg Tube Maximum Radius Tube Adapter No x Vol ml Description RPM RCF Quantity x Cat No 24 x 1 5 2 0 mL 11x54 mm microtube 15000 21000 8 35 24 x 0 5 mL 8 x 54 mm microtube 15000 17432 6 93 24 x 0 5 mL 8 x 54 mm microtube 15000 18740 7 45 2 pr x 5763E 24 x 0 4 mL 6 x 50 mm microtube 15000 21000 8 35 2 pr x 5764E 24 x 0 25 mL 6 x 50 mm microtube 15000 18240 7 25 2 pr x 5764E Tube M...

Страница 14: ...700 RPM 4 000 x 925 3 700 Specific gravities greater than 3 0 This table is based on the formula Use the same formula to compute derating factors for specific gravities greater than 3 0 S0 is the maximum specific gravity allowed before derating 1 2 Sa is the actual specific gravity of the sample CAUTION Do not exceed the rated speed or specific gravity Higher speeds or specific gravities will impo...

Страница 15: ...irs For Oxidizing Acids see Oxidizing Agents strong Derating Table for Microplate Rotor Load per Carrier grams Max Speed rpm 320 3650 340 3550 360 3450 380 3400 400 3300 450 3100 500 2950 600 2700 Greater than 600 do not use Plastic Metal Other POLYALLOMER POLYCARBONATE POLYETHYLENE POLYPROPYLENE POLYURETHANE MODIFIED PHENYLENE OXIDE NORYL ACETAL HOMOPOLYMER DELRIN ACETAL COPOLYMER CELCON NYLON PO...

Страница 16: ...EL ALUMINUM MAGANESE BRONZE MAGNESIUM RUBBER BUNA N VITON PHENOLIC FIBER PT PAINTED SURFACE Mechanical Autoclave S M U S M U S S S U S S S S S S S M S M Ethylene Oxide Gas S S S S S S S S S S S S S S S U U S S S Dry Heat 2Hrs 160 C U U U U U U U U U U S S U S S U U U U U Chemical Ethanol S S S S U S S S U M S S S S S S S S S S 40 Formalin S S S S U S S S S U S S S S S S U S S S Methanol S M S S M ...

Страница 17: ...OM6464K Rev 4 11 January 2004 Page 17 of 24 4 6 RCF Nomograph ...

Страница 18: ... remove the corrosion discontinue use of the product and inform technical service 5 3 Storage Keep the Unit Dry Store parts on a soft surface to avoid damage Rotors and other parts should be clean and dry Store them open to the air not in a plastic bag so any residual moisture evaporates Face the parts upward to avoid moisture retention in the cavities 5 4 Decontamination Procedures WARNING WARNIN...

Страница 19: ...y 9 to 40 C by 1 Rotation Counterclockwise RPM 500 15 000 by 50 RPM RCF 50 21 000xg by 50 xg Spin Duration 0 99 min 59 sec by 1 min and or 1 sec HOLD MODE Momentary Spin Brake Active Electric 3 profiles Repeatability Temperature Control IEC Centra CL5R only 2 C within range of 4 C to ambient Rotation Accuracy within 10 RPM Timer Microprocessor controlled Accuracy within 1 sec System Components Mot...

Страница 20: ...OM6464K Rev 4 11 January 2004 Page 20 of 24 ...

Страница 21: ...ANY PERSON OR ENTITY OTHER THAN THERMO WITHOUT THERMO S PRIOR WRITTEN APPROVAL OR ANY USE OF REPLACEMENT PARTS NOT SUPPLIED BY THERMO SHALL IMMEDIATELY VOID AND CANCEL ALL WARRANTIES WITH RESPECT TO THE AFFECTED PRODUCTS THE OBLIGATIONS CREATED BY THIS SECTION TO REPAIR OR REPLACE A NON CONFORMING PRODUCT SHALL BE THE SOLE REMEDY OF PURCHASER IN THE EVENT OF A NON CONFORMING PRODUCT EXCEPT AS EXPR...

Страница 22: ...OM6464K Rev 4 11 January 2004 Page 22 of 24 Notes ...

Страница 23: ...OM6464K Rev 4 11 January 2004 Page 23 of 24 ...

Страница 24: ...2778 1101 Fax 91 22 2778 1103 Bangaalore India 81 45 453 9122 Fax 81 45 453 9222 Yokohama City Japan International Sales Office Locations Bioscience Technologies Thermo Electron Corporation 866 9 THERMO 866 984 3766 Molecular Biology info molbio thermo com Sample Preparation info sampleprep thermo com New Labs newlabs thermo com Services services biosciencetech thermo com Laboratory Pipetting and ...

Отзывы: