5.3  Replacing the Temperature Sensor

1. Locate the probe mounting plate in the center of the

right side of the incubator interior.

2. Open the mounting plate by removing the screws that

hold it in place.

3. Locate the sensor mounted on the inside of the panel in

a black housing. Note the angle of the probe.

4. Grasp and unplug the probe from the probe cable.
5. When replacing the sensor, be sure to mount the probe at

the same angle as it was originally mounted.

5.4  Removing the Top Section

If it becomes necessary to remove the top of the incubator

for moving it through low doorways, refer to the points follow-
ing, then use the indicated procedure:  

• Read the instructions completely before starting the

removal process. 

• Provide adequate space and sufficient lighting to perform

the work.

• Disconnect the inlet water supply and the drain.
• Before lifting the top section from the incubator, provide

carpenter’s horses or another suitable support arrangement
so that the underside of the top is suspended. 

Procedure:

1. Remove all test material from the incubator chamber.
2. All temperature-sensing probes are mounted on a probe-

mounting bracket on the right side of the incubator
chamber.  Remove the nuts securing the bracket and
then remove the probes from the bracket. Note the loca-
tion and configuration of the probes.

3. The sensing bulb capillaries are routed behind the duct

sheet, up to the access port at the top of the unit and
behind the control panel.

Note:

Route all the probes between the duct sheet and the out-

side wall of the chamber so that when the top is raised, all
probes will move freely up and out. Do not kink or bend the
capillaries. 

4. Open the control panel by  removing the two screws

located on the left side of the control panel.

5. Loosen the top gasket around each of the air exhaust

vents by turning the screw.

6. Remove the eight screws from the top cover of the incu-

bator.

7. Remove the four screws from each of the vent caps on

the top cover of the incubator. The top cover can now be
removed.

Section 5 - Service

Servicing must be performed by qualified service
personnel only!  

De-energize all potential sources of energy to this
unit and lockout/tagout their controls. (O.S.H.A.
Regulation, Section 1910-147.)  

5.1  Access to the Electrical Components

To gain access to the electrical components, remove the

two screws located on the left side of the control panel with a
phillips screwdriver. The control panel is hinged and will swing
open.

5.2  Replacing the Overtemp/Undertemp Probe and
Thermostat

1. Remove the incubator ceiling by remove the screws

holding it in place.

2. Remove the top three screws from the top of the right

duct cover.

3. Lean the duct sheet out, and remove the Permagum seal

from around the probe access hole.

4. Remove the 15” copper capillary overtemp probe by

extracting the two plastic clips that hold the probe in
place.

5. Open the control panel by removing the two screws

located on the left side of the control panel.

6. Pull the overtemp probe up through the access hole and

into the control panel.

7. Follow the wires from the probe to the thermostat

mounted on the control panel. Clip the plastic ties hold-
ing the overtemp cable to the existing wiring.

8. Pull the overtemp knob on the control panel off.
9. Remove the two screws that hold the overtemp assembly

to the control panel.

10. Disconnect the two wires from the back of the thermo-

stat assembly.

11. Pull the entire assembly out of the panel, and remove the

unit.

12. Replace the thermostat and probe.

Note:

Reseal the probe access hole with Permagum, and retie

the overtemp cable to the existing wires after replacing the
probe.

Model 3920 _______________________________________________________________________________Service

5 - 1

Содержание 3920

Страница 1: ...Model 3920 29 cu ft Capacity Environmental Chamber Operating and Maintenance Manual Manual No 7033920 Rev 15 ...

Страница 2: ...MANUAL NUMBER 7033920 15 21505 SI 9018 9 18 03 New drier P N 209016 no flare connections ccs 14 21173 IN 3057 11 15 02 Added ground shields on temperature and humidity probes ccs 13 20479 IN 3013 6 27 02 Removed Auto IR CO2 option as a standard accessory aks 12 20479 IN 3013 5 29 02 Added Auto IR CO2 option as a standard accessory aks 20480 PIP 041 11 21 01 Removed reference to p trap 190467 no lo...

Страница 3: ...mientos asociadas con este simbulo Equipment being maintained or serviced must be turned off and locked off to prevent possible injury Risques potentiels liés à l énergie L équipement en entretien ou en maintenance doit être éteint et mis sous clé pour éviter des blessures possibles Geräte an denen Wartungs oder Servicearbeiten durchgeführt werden müssen abgeschaltet und abgeschlossen wer den um V...

Страница 4: ...Model 3920 _________________________________________________________________________________ iii ...

Страница 5: ...b Changing the Chart Paper 2 3 c Changing the Pen 2 3 d Calibrating the Chart Recorder 2 3 2 9 Offset Calibration 2 4 2 10 Honeywell Recorder 2 4 2 11 Controller Configuration 2 4 Section 3 Control Panel Operation 3 1 3 1 The 3920 Control Panel 3 1 3 2 Setting the Operating Temperature Figure 3 3 3 2 3 3 Programming the controller 3 2 a Removing the mechanical lockout 3 2 b Removing the Software L...

Страница 6: ... personnel sign a free and clear delivery receipt Whenever possible Thermo will assist in settling claims for loss or in transit damage Section 2 Installation and Start Up 2 1 Location Locate the unit on a firm level surface in an area of mini mum ambient temperature fluctuation A minimum of 12 inches clearance is required at the top of the incubator and a minimum of 3 inches clearance is required...

Страница 7: ...ise approximately two units on the scale The alarm should be silenced and the overtemp indicator light should go out The overtemp safety thermostat is now set a few degrees above the control temperature setpoint When the chamber tem perature rises to the overtemp control point the alarm system will activate power to the heaters will shut off and the cham ber temperature will be maintained at the o...

Страница 8: ...and arm as an assembly 2 Unsnap the plastic hinge securing the pen Remove and discard the old pen 3 Install the new pen by snapping the hinge securely around the pen arm 4 Re install the pen assembly by sliding the pen arm under the screws positioning the pen tip in the time line groove Tighten the screws 5 Push the Chart Change button Figure 2 4 and hold it for 1 second until the pen begins to mo...

Страница 9: ...the controller and the reference sensor value 12 Press the Display key once Model 3920__________________________________________________________________Installation and Start Up Press Press Side View To turn the hardware lockout back on 1 Press the Up and Down Arrow keys simultaneously for 3 seconds The word InPt will appear in the upper dis play and set will appear in the lower display 2 Press th...

Страница 10: ... Ct1 5 rA2 0 25 It1 NA Pb2 1 0 dE2 NA rA1 0 19 rE2 0 25 Ct2 30 PROG See Programming Sheets if required Setup Menus InPt In1 RT D CAL1 rL2 NA CAL2 NA dEC1 NA rtd1 DIN rH2 NA Hunt NA rL1 20 0 Ftr1 2 LrnL NA rHi 80 0 In2 NA LrnH NA OtPt Ot1 Ht AL2 NA LAt3 nLA SIL4 NA PrC1 NA LAt2 NA SIL3 OFF Aout NA HyS1 0 1 SIL2 NA Ot4 NA PrC3 NA Ot2 CL Ot3 AL3 AL4 NA ArL NA PrC2 NA AL3 Pr1 HyS4 NA ArH NA HyS2 0 1 H...

Страница 11: ...te the audible alarm and the overtemp indicator light Interrupt power to the heaters and maintain the incubator s cabinet temperature at the overtemp safety control point Note The overtemp control is not directly calibrated The numbers 0 to 10 on the dial are for reference only and do not correspond to any specific temperature If an overtemp condition occurs the alarm can only be silenced by raisi...

Страница 12: ...ckout Make sure the unit is turned off Wearing a grounding wrist strap or maintaining constant contact with the metal cabinet press in the four locking tabs on the frame of the Temperature controller There are two tabs on either side as shown in the front and side views in Figure 3 4 When all tabs are unlocked pull the con troller module out of its frame Looking at the top of the module locate the...

Страница 13: ...ormance of the unit the ventilator caps should be closed at all times Figure 3 9 Section 4 Routine Maintenance 4 1 Cleaning the Incubator De energize all potential sources of energy to this unit and lockout tagout their controls O S H A Regulation Section 1910 147 The continued cleanliness of the stainless steel used in our products has a direct effect on the appearance and operation of the unit U...

Страница 14: ...d list of preventive maintenance requirements See the specified section of the instruction manual for further details We have qualified service technicians using NIST traceable instruments available in many areas For more information on Preventive Maintenance or Extended Warranties please contact us at the number below Cleaning and calibration adjustment intervals are dependent upon use environmen...

Страница 15: ...aps for adjustment open or close as required 4 1 Perform a complete decontamination procedure Wipe down interior shelves Between Experiments side panels with disinfectant Rinse everything well with sterile distilled water More frequent decontamination may be required depending on use and environmental conditions 2 9 Verify and document all calibrations at the minimum Clean drip pan and drain lines...

Страница 16: ...top gasket around each of the air exhaust vents by turning the screw 6 Remove the eight screws from the top cover of the incu bator 7 Remove the four screws from each of the vent caps on the top cover of the incubator The top cover can now be removed Section 5 Service Servicing must be performed by qualified service personnel only De energize all potential sources of energy to this unit and lockou...

Страница 17: ...move the black trim gasket located at the junction of the top assembly and the main incubator section The ends of the gasket have been joined together at the rear of the incubator Note When raising the top section note the orientation of the gasket seal at the top opening of the incubator chamber The gasket must be correctly positioned when reinstalling the top on the chamber 10 Slowly lift the en...

Страница 18: ...ompression Finish Powder coated Salt spray tests exceed 1000 hrs per ASTM Standard B117 85 Weights Net 700 lbs Shipping Motor 850 lbs Temperature Alarm Sensor Thermostat Controller Thermostat Setpoint Analog reference dial Alarm Audible visual Fittings Drain Port 3 8 OD Copper Unit Heat Load 115V 5500 BTUH 1600W 220V 6000 BTUH 1750W Refrigeration Compressor 1 4 Horsepower air cooled R 134A Electri...

Страница 19: ...20277 Triac 25A 2 403940 Thermostat B 10 0 220 F coated 1 403941 Thermostat B 20 10 220 F coated 1 131300 Blower Wheel 1 275030 Defrost Timer 1 630090 Heater 2 205062 Condensing Unit 1 4 HP 115V R 134A 1 991334 Solenoid 120V 101 Port 1 285599 Thermal Fuse 250V 17A 1 830044 Blower Motor 1 4 HP 115V 1 Immediate Stock Model 3920 ________________________________________________________________________...

Страница 20: ...Model 3920 _________________________________________________________________Refrigeration Schematics 8 1 ...

Страница 21: ...Model 3920 ____________________________________________________________________Electrical Schematics 8 2 ...

Страница 22: ...Model 3920 ____________________________________________________________________Electrical Schematics 8 3 ...

Страница 23: ...Model 3920 ____________________________________________________________________Electrical Schematics 8 4 ...

Страница 24: ...Model 3920 ____________________________________________________________________Electrical Schematics 8 5 ...

Страница 25: ...Thermo Electron Corporation Controlled Environment Equipment Millcreek Road P O Box 649 Marietta Ohio 45750 U S A Telephone 740 373 4763 Telefax 740 373 4189 ...

Отзывы:

Похожие инструкции для 3920