background image

 
 
 
 

 

 

 

 

 

Pierce Biotechnology 

PO Box 117 

(815) 968-0747 

thermofisher.com 

 

 

 

3747 N. Meridian Road 

Rockford, lL 61105 USA 

(815) 968-7316 fax 

 

 

Procedure for Desalting or Buffer Exchange 

Additional Materials Required 

 

For 75µL and 0.5mL spin columns: Bench-top microcentrifuge (1500 

×

 

g

) and 1.5mL microcentrifuge tubes 

Note:

 Use a centrifuge that can be adjusted to 1500 

×

 

g

, such as the Thermo Scientific™ Sorvall™ Legend Micro 17 

Microcentrifuge 

 

For 2 and 5mL spin columns: Centrifuge (1000 

×

 

g

) and 15mL conical tubes  

 

For 10mL spin columns: Centrifuge (1000 

×

 

g

) and 50mL conical tubes  

 

For the desalting plates: Variable-speed centrifuge with rotor and carrier capable of handling stacked plates                        
(height = 4.4cm) at 1000 

×

 

g

 

 

Wash/equilibration buffer  

Note:

 Use the same wash/equilibration (stacker) buffer as is desired for the final sample solution. Equilibrating the 

desalting resin before sample loading is necessary to ensure proper buffer exchange.  

 

Procedure for Protein Desalting  

Note:

 See 

Table 1

 for centrifugation times and volumes for the buffer, stacker and sample for each device. For maximum 

protein recovery, add a stacker on top of the applied sample for volumes below the volume specified in 

Table 1

.  

1.

 

Remove the column’s bottom closure or the plate’s bottom sealing material. Loosen cap (do not remove cap).  

2.

 

Place the column into a collection tube or plate on top of a wash plate and centrifuge to remove the storage solution. 

3.

 

Discard flow-through and replace the column back into the collection device. 

4.

 

Add wash/equilibration buffer on top of the resin. Centrifuge device and discard flow-through. Repeat this step two 
additional times. 

Note:

 After each spin, the resin should appear white and free of liquid. If liquid is present, make sure you are using the 

correct centrifugation speed and time. Incomplete centrifugation may result in poor sample recovery or sample dilution.  

5.

 

Blot the bottom of the column or plate to remove excess liquid. Transfer device to a new collection tube or plate.  

6.

 

Apply sample on top of the resin. If needed, add a stacker as soon as the sample has entered the resin. Adding a stacker is 
optional but recommended for dilute protein solutions or small sample volumes to ensure maximum sample recovery. 

7.

 

Centrifuge and retain flow-through that contains sample. Discard spin column or plate. 

 

Table 1. Centrifugation times and volumes for the buffer, stacker and sample.

 

Column or Plate

 

75µL 

0.5mL 

2mL 

5mL 

10mL 

Plate 

Sample Volume Range (

µ

L) 

2-12 

30-130 

200-700 

500-2000 

700-4000 

20-100 

Wash/equilibration Buffer Volume 

50µL 

300µL 

1mL 

2.5mL 

5mL 

250µL 

Sample Volume (µL)* 

< 5 

< 70 

< 350 

< 750 

< 1500 

< 30 

Optional Stacker Volume (µL)* 

15 

40 

100 

200 

10 

Centrifuge Speed (

×

 g

1000 

1500 

1000 

1000 

1000 

1000 

Centrifugation 

Time (min) 

Storage Solution 

Removal 

Wash 1 

Wash 2 

Wash 3 

Sample Recovery 

*When using the indicated sample volumes, use a stacker to achieve the highest recovery. The stacker is a volume 
of wash/equilibration buffer applied after the added sample has completely entered the desalting resin bed. 

 

Отзывы: