Troubleshooting 

Problem 

Possible Cause 

Solution 

Sample or buffer does not 
flow through resin 

Centrifugation problem 

Ensure that centrifuge is in proper working condition 

Ensure bottom closure is removed 

Ensure top cap is loosened 

Sample contamination 

Improper sample loading 

Apply sample directly to center of the resin bed; touch tip 
to resin to expel all sample 

Avoid contact with sides of column 

Improper centrifugation  

For fixed-angle rotors, place column in the same orientation 
each time and do not exceed recommended centrifuge speed 

Do not exceed recommended centrifugation speed or time 

Low yield 

Sample was not completely in 
solution before adding to the 
column 

Centrifuge sample at 14,000 

×

 

g

 for 10 minutes before 

adding to the column 

Portion of protein still 
remaining in spin column 

Use a stacker to recover more protein (for most samples, 
the majority of protein is recovered without a stacker) 

Protein precipitated in 
equilibration buffer  

Check for protein solubility in the final buffer or solution 

Protein bound to resin matrix 

Use alternative resin, such as Polyacrylamide Spin 
Desalting Columns (see Related Thermo Scientific Products 
section) 

Recovered protein or 
sample is dilute 

Wash/equilibration buffer was 
not adequately removed 

Before adding the sample make sure the wash/equilibration 
buffer was adequately removed by centrifugation (i.e., 
column appears uniformly white with no solvent streaks) 

 

 
Related Thermo Scientific Products 

89849 

Pierce

 

Polyacrylamide Spin Desalting Columns, 7K MWCO, 0.7mL, 25 columns 

89862 

Pierce

 

Polyacrylamide Spin Desalting Columns, 7K MWCO, 0.7mL, 50 columns 

 
 

Products are warranted to operate or perform substantially in conformance with published Product specifications in effect at the time of sale, as set forth in 
the Product documentation, specifications and/or accompanying package inserts (“Documentation”). No claim of suitability for use in applications regulated 
by FDA is made. The warranty provided herein is valid only when used by properly trained individuals. Unless otherwise stated in the Documentation, this 
warranty is limited to one year from date of shipment when the Product is subjected to normal, proper and intended usage. This warranty does not extend to 
anyone other than Buyer. Any model or sample furnished to Buyer is merely illustrative of the general type and quality of goods and does not represent that 
any Product will conform to such model or sample. 
NO OTHER WARRANTIES, EXPRESS OR IMPLIED, ARE GRANTED, INCLUDING WITHOUT LIMITATION, IMPLIED WARRANTIES OF 
MERCHANTABILITY, FITNESS FOR ANY PARTICULAR PURPOSE, OR NON INFRINGEMENT. BUYER’S EXCLUSIVE REMEDY FOR NON-
CONFORMING PRODUCTS DURING THE WARRANTY PERIOD IS LIMITED TO REPAIR, REPLACEMENT OF OR REFUND FOR THE NON-
CONFORMING PRODUCT(S) AT SELLER’S SOLE OPTION. THERE IS NO OBLIGATION TO REPAIR, REPLACE OR REFUND FOR PRODUCTS 
AS THE RESULT OF (I) ACCIDENT, DISASTER OR EVENT OF FORCE MAJEURE, (II) MISUSE, FAULT OR NEGLIGENCE OF OR BY BUYER, 
(III) USE OF THE PRODUCTS IN A MANNER FOR WHICH THEY WERE NOT DESIGNED, OR (IV) IMPROPER STORAGE AND HANDLING OF 
THE PRODUCTS. 

Unless otherwise expressly stated on the Product or in the documentation accompanying the Product, the Product is intended for research only and is not to 
be used for any other purpose, including without limitation, unauthorized commercial uses, in vitro diagnostic uses, ex vivo or in vivo therapeutic uses, or 
any type of consumption by or application to humans or animals. 

Current product instructions are available at 

thermofisher.com

. For a faxed copy, call 800-874-3723 or contact your local distributor. 

© 2016 Thermo Fisher Scientific Inc. All rights reserved. Unless otherwise indicated, all trademarks are the property of Thermo Fisher Scientific Inc. and its 
subsidiaries. Printed in the USA. 

 

Отзывы: