Troubleshooting
Problem
Possible Cause
Solution
Sample or buffer does not
flow through resin
Centrifugation problem
Ensure that centrifuge is in proper working condition
Ensure bottom closure is removed
Ensure top cap is loosened
Sample contamination
Improper sample loading
Apply sample directly to center of the resin bed; touch tip
to resin to expel all sample
Avoid contact with sides of column
Improper centrifugation
For fixed-angle rotors, place column in the same orientation
each time and do not exceed recommended centrifuge speed
Do not exceed recommended centrifugation speed or time
Low yield
Sample was not completely in
solution before adding to the
column
Centrifuge sample at 14,000
×
g
for 10 minutes before
adding to the column
Portion of protein still
remaining in spin column
Use a stacker to recover more protein (for most samples,
the majority of protein is recovered without a stacker)
Protein precipitated in
equilibration buffer
Check for protein solubility in the final buffer or solution
Protein bound to resin matrix
Use alternative resin, such as Polyacrylamide Spin
Desalting Columns (see Related Thermo Scientific Products
section)
Recovered protein or
sample is dilute
Wash/equilibration buffer was
not adequately removed
Before adding the sample make sure the wash/equilibration
buffer was adequately removed by centrifugation (i.e.,
column appears uniformly white with no solvent streaks)
Related Thermo Scientific Products
89849
Pierce
Polyacrylamide Spin Desalting Columns, 7K MWCO, 0.7mL, 25 columns
89862
Pierce
Polyacrylamide Spin Desalting Columns, 7K MWCO, 0.7mL, 50 columns
Products are warranted to operate or perform substantially in conformance with published Product specifications in effect at the time of sale, as set forth in
the Product documentation, specifications and/or accompanying package inserts (“Documentation”). No claim of suitability for use in applications regulated
by FDA is made. The warranty provided herein is valid only when used by properly trained individuals. Unless otherwise stated in the Documentation, this
warranty is limited to one year from date of shipment when the Product is subjected to normal, proper and intended usage. This warranty does not extend to
anyone other than Buyer. Any model or sample furnished to Buyer is merely illustrative of the general type and quality of goods and does not represent that
any Product will conform to such model or sample.
NO OTHER WARRANTIES, EXPRESS OR IMPLIED, ARE GRANTED, INCLUDING WITHOUT LIMITATION, IMPLIED WARRANTIES OF
MERCHANTABILITY, FITNESS FOR ANY PARTICULAR PURPOSE, OR NON INFRINGEMENT. BUYER’S EXCLUSIVE REMEDY FOR NON-
CONFORMING PRODUCTS DURING THE WARRANTY PERIOD IS LIMITED TO REPAIR, REPLACEMENT OF OR REFUND FOR THE NON-
CONFORMING PRODUCT(S) AT SELLER’S SOLE OPTION. THERE IS NO OBLIGATION TO REPAIR, REPLACE OR REFUND FOR PRODUCTS
AS THE RESULT OF (I) ACCIDENT, DISASTER OR EVENT OF FORCE MAJEURE, (II) MISUSE, FAULT OR NEGLIGENCE OF OR BY BUYER,
(III) USE OF THE PRODUCTS IN A MANNER FOR WHICH THEY WERE NOT DESIGNED, OR (IV) IMPROPER STORAGE AND HANDLING OF
THE PRODUCTS.
Unless otherwise expressly stated on the Product or in the documentation accompanying the Product, the Product is intended for research only and is not to
be used for any other purpose, including without limitation, unauthorized commercial uses, in vitro diagnostic uses, ex vivo or in vivo therapeutic uses, or
any type of consumption by or application to humans or animals.
Current product instructions are available at
thermofisher.com
. For a faxed copy, call 800-874-3723 or contact your local distributor.
© 2016 Thermo Fisher Scientific Inc. All rights reserved. Unless otherwise indicated, all trademarks are the property of Thermo Fisher Scientific Inc. and its
subsidiaries. Printed in the USA.