A
Using the Direct Sample Probe
Running the Sequence
Thermo Scientific
TSQ Quantum XLS Series User Guide
137
d. Click any
Yes
in the In Use column to clear all instruments other than the TSQ
Quantum.
e.
Ensure that the TSQ Quantum has a Yes in the In Use column. Click the area if
necessary.
f.
Click any
Yes
in the Start Instrument column to clear all start instruments.
g. Click
OK
to save your settings and close the Change Instruments In Use dialog box.
Running the Sequence
You acquire data by running the sequence. The Xcalibur data system saves the acquired data
in a raw file.
Y
To run the sequence
1. In the Run Sequence dialog box (see
on
OK
to start the
acquisition. The Xcalibur system displays a message to insert the probe (
Figure 67.
Insert probe message
2. Inset the probe into the inlet valve:
a. Insert the guide ball on the DIP probe into the guide ball hole on the guide bar.
b. Slide the probe forward into the inlet valve until the guide ball is at the guide bar’s
first stop. See
.
c.
Tighten the inlet valve knob clockwise to ensure a leak-tight seal.
Note
You must have the TSQ EZ Tune or QuickQuan application running so that the
Xcalibur application can display instrument methods.
Содержание TSQ Quantum XLS
Страница 1: ...TSQ Series TSQ Quantum XLS Series User Guide 70111 97225 Revision A December 2011 ...
Страница 3: ......
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 32: ......
Страница 64: ......
Страница 76: ......
Страница 102: ...86 TSQ Quantum XLS Series User Guide Thermo Scientific Figure 50 Ion source assembly assembled and exploded views ...
Страница 160: ......
Страница 166: ......
Страница 176: ......