3
Advanced Troubleshooting
Replacing the Collision Gas Components
382
TSQ 9000 Hardware Manual
Thermo Scientific
Replacing the Collision Gas Components
The collision gas components you can order and replace in the TSQ 9000 system are the
collision gas module, the collision gas tubing, the tee that connects the collision gas tubing to
the foreline adapter, and the foreline adapter itself.
To replace the collision gas module, collision gas tubing, tee, and foreline adapter
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Slide the top cover panel off the instrument.
3. Slide the top cover panel toward the back of the instrument and lift it off.
4. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 407.
Locating the Left Hand Front Panel Screws
5. Lift the left hand front panel off the instrument.
6. Open the front door of the GC and disconnect the column and transfer line nut.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Screws
Left Hand Front
Panel
Screw on Front
of Instrument