Medical Devices -Touch Controller

9

 |   Transport and Handling

Transport and Handling

The machine must be transported and handled solely in 
a vertical position and following the instructions printed 
on the packaging. This precaution is necessary to avoid 

contamination of the compressor with oil which may 
cause the rupture of valves of the cooling coils and 
problems with electric motor starting.
The accessories supplied with the machine (slides, wire 

shelves, drawers, baskets, etc.) are shipped inside the 
unit. The machine is fixed on a wooden pallet by means 
of screws, wrapped with polyethylene and packaged in 

cardboard, a wooden crate or wooden case.
The machine must be handled using a forklift or pallet 
truck with suitable forks (fork length at least equal to 
2/3 length of the unit).

Positioning

Incorrect positioning can cause damage to the machine 
and create dangerous conditions for users; therefore, 

the installer must comply with the following general 
rules:

Figure 1. Equipment Position

-

Set the machine in a sufficiently ventilated 
environment.

-

Place the machine far from heat sources and far 

from sources of electromagnetic interferences 
(such as motors, generators, infrared beams, 
telephones) which can have negative effects on 

equipment functioning.

-

Avoid exposure to direct sunlight and air 
conditioning flows.

-

Remove the supplied accessories and the wooden 
pallet base.

-

Position the machine with the aid of a spirit level. 

If the machine needs to be set down in 
order to bring it into the installation location 
it is necessary to wait at least 6 hours 

before switching it on.
The manufacturer declines any 
responsibility for problems due to 
transport carried out in any condition 

different from those specified above.

Position the machine keeping a minimum 
distance of 100 mm from all walls. If the 
machine is embedded in any type of 

furniture, a correct air flow of the 
condensing unit (compressor/fan motors) 
must be always ensured. The warranty will 

immediately expire, if this is not 
guaranteed. (Refer to 

Figure 1

).

For models taller than 1.5 m, installation 
with wall fixing brackets is recommended.

100 mm

100 

100 

mm

100 

100 

mm

100

m

100

m

mm

100

m

100

m

mm

100 mm

Содержание TSB140F

Страница 1: ...ww thermofisher com usermanuals pour obtenir des manuels en fran ais en allemande en italien et en espagnol Visitare www thermofisher com usermanuals per i manuali in francese Tedesco italiano e spagn...

Страница 2: ...nuel recherch veuillez contacter le service clients Vous trouverez les coordonn es sur thermofisher com contactus Visitare www thermofisher com usermanuals per i manuali in francese tedesco italiano e...

Страница 3: ...nd Warranty 5 Purpose Content and Recipients of the Manual 5 Arrangements Prepared by the Customer 6 Product Description 7 Technical Description 7 Intended Use 7 General Device Features 8 Climatic Cla...

Страница 4: ...or 37 Ordinary and Extraordinary Maintenance 39 Prohibition of Safety Device Removal 39 Cleaning the Unit Inside and Outside 39 Cleaning the Condenser 39 Condensate Water Draining 41 Replacing Buffer...

Страница 5: ...Models Models Model Temperature at Factory Set point TSB140R 2 C to 8 C TSB250R 2 C to 8 C TSB400R 2 C to 8 C TSB700R 2 C to 8 C TSB1500R 2 C to 8 C TSB140F 12 C to 25 C TSB250F 30 C to 40 C TSB400F...

Страница 6: ...ntenance operations must be carried out by personnel authorized by the manufacturer The corresponding warranty will become immediately void in the event of use or maintenance of equipment not in compl...

Страница 7: ...s Noise This machine has been designed and built to minimize risks related to the emission of acoustic noise always lower than 70 dB Contraindications The refrigerated cabinet must not be used Exposed...

Страница 8: ...78 the environment must have a volume of 1m3 every 8gr of R290 refrigerant The quantity of gas contained in the circuit is declared on the silver data plate attached inside the cabinet 2 Do not use th...

Страница 9: ...to the repair or replacement of goods with the absolute exclusion of any direct or indirect damages of any kind In any case the right to repair or replacement of materials must be exercised within th...

Страница 10: ...ouch Controller Arrangements Prepared by the Customer The following arrangements are set by the customer The machine electrical connection Installation site arrangement Routine maintenance Refrigerato...

Страница 11: ...d on all Thermo Fisher Scientific models is 30 kg with weight uniformly distributed Intended Use All the listed models are suitable for storage For this reason we suggest only storing products that ha...

Страница 12: ...gether It is important to allocate devices to store one or the other Avoid mixed use IMPORTANT Manufacturer declines responsibility in case staff in charge is not adequately formed and trained on bloo...

Страница 13: ...sition Set the machine in a sufficiently ventilated environment Place the machine far from heat sources and far from sources of electromagnetic interferences such as motors generators infrared beams t...

Страница 14: ...into alternating current or three phase current or energy saving connector Could cause damage to the electronics The glass door must be cleaned using a cloth dampened with water Do not therefore use...

Страница 15: ...ck system allows for the installation of fully extractable drawers on telescopic guides with bayonet connection Drawers Installation Insert the guides into the special slots of racks Figure 4 and pull...

Страница 16: ...place material over the label indicating the maximum permitted loading level if present Store material leaving a distance of at least 60 mm from sides and at least 180 mm from the top of the compartm...

Страница 17: ...nt air vents Load stored material gradually at ambient temperature to grant proper refrigeration Store material on drawers Do not place products directly on the bottom or against the wall doors or fix...

Страница 18: ...itching On To switch on the equipment for the first time follow these below instructions Table 2 Instructions for switching ON Sl No Instructions Graphical Representation 1 Connect the plug to the ele...

Страница 19: ...ct language and set date and hour 4 User interface will be displayed Equipment with one chamber Table 2 Instructions for switching ON Sl No Instructions Graphical Representation Temperature detected i...

Страница 20: ...al Devices Touch Controller Figure 12 Equipment with One Chamber Ballasted Temperature Product Probe Figure 13 Ballasted Temperature Product Probe User Hardware Interface Figure 14 Hardware Interface...

Страница 21: ...set back alarms and failures Working Status N D Function Table 3 Working Status Strings STRING WORKING STATUS PAUSE Compressor is off waiting for next cooling cycle COOLING The compressor is ON to re...

Страница 22: ...ction is activated see Night Day Function Setpoint Adjustment Limits Setpoint can be adjusted only according to relative guidelines for storage and preservation of blood plasma and breast milk Every t...

Страница 23: ...ouch lower or upper limit to access to the dedicated page Touch NEW DOOR DELAY and enter the new value Press OK to confirm it Go back to homepage by touching home icon Duration of storage is strictly...

Страница 24: ...battery backup installed Go back to homepage by touching home icon Electric Digital Key Lock As safety device against unauthorized door opening the equipment can be fit with an electric digital key lo...

Страница 25: ...tored data flash memory stores up to 1 year in case SD card is not in its slot it has been lost or damaged SD Info Date of last backup No Files included Free space No of date hour changes done by the...

Страница 26: ...32 days in which door opening alarm has occurred It allows access to the night day function and to the key locking management settings key locking is password protected if password is enabled It runs...

Страница 27: ...ction is enabled The display logo and led light brightness is lowered LED light inside the chamber will turn blue The temperature inside the chamber is increased of a default value from 0 1 C to 2 C m...

Страница 28: ...n accordance with its optimal use In addition to the standard opening by touching the relating icon possible configurations are nr 2 Lock always open touch to activate Automatic opening and closing ac...

Страница 29: ...ioning Despite this in user panel the defrost icon is available to run manual defrost if necessary and only under specific conditions Multimedia Information TUTORIAL The multimedia section enables the...

Страница 30: ...ight happen when civil time is changed to solar time and vice versa there might be data loss while downloading data from USB Therefore we advise to backup data from SD CARD see Temperature Data Backup...

Страница 31: ...gement from the SYSTEM SETUP menu This user has no access credentials and he can only modify the password By default the password for managing the CONTROL user is 0000 But by accessing this menu and s...

Страница 32: ...or opening the date hour changing etc ScreenSaver User can enable screensaver entering software panel setup Screensaver will automatically activate if the screen is not in use for a certain period of...

Страница 33: ...imit set Once it happens acoustic signalling takes place three times and dry contact switches To conclude test the value of the probe turns back close to the device operating value after that it is al...

Страница 34: ...icon of alarm will appear littler and near the temperature of the chamber in order to signal that the alarm is still running If after 30 minutes the Failure Alarm persists the acoustic alarm will be...

Страница 35: ...temperature backup folder Docs including Capture 5 2 software manuals and electric drawings Recovery including the backup of all the customization made by the user Follow the below steps to install Ca...

Страница 36: ...is possible to visualize the temperature graph of the next day You can build a weekly chart by clicking Clicking again the next week temperature chart will be visualized The scale on the axis of the g...

Страница 37: ...uch Controller 33 Touchscreen Controller Display Functioning How to Export Data Click and save the file for example save it with equipment serial number Open Excel software click Data and load the fil...

Страница 38: ...34 Touchscreen Controller Display Functioning Medical Devices Touch Controller Then follow these steps...

Страница 39: ...Medical Devices Touch Controller 35 Touchscreen Controller Display Functioning This is the exported table...

Страница 40: ...e setup according to your corporate network so follow these steps In the homepage touch first user menu icon and second service menu icon Now the service menu password is asked Enter 2810 PSW and conf...

Страница 41: ...ith 4 20mA input 4 20mA outputs convert a temperature value detected by the touchscreen controller or DMLP digital monitor probe into an electric signal Manual Unlocking of the Door In the event of un...

Страница 42: ...door manually in equipment with one chamber For model 1500 manual door opening has to be performed by accessing from top of the refrigerator Figure 24 Pulling the release lever upwards and keeping le...

Страница 43: ...d Cleaning products Water and non abrasive neutral detergents Do not use solvents or thinners Cleaning method Use a cloth or sponge soaked in a suitable cleaning product to clean the inside and outsid...

Страница 44: ...ore the electrical current and switch back on the device Figure 26 Condenser Position in models 140 MODEL 250 400 PHASE 1 Use a Phillips head screwdriver to remove the guard 3 screws PHASE 2 Use a vac...

Страница 45: ...ith motor compartment in lower position In some other models water is collected in a tray included in supply which is set under the unit and inserted in the corresponding slides This tray must be empt...

Страница 46: ...es To ensure maximum efficiency we suggest periodically changing backup batteries at least every 2 years This operation can be performed by users easily accessing the housing located at the rear of th...

Страница 47: ...electronic appliances Before scrapping the machine make it unusable by cutting the power cord and removing the doors shelves and drawers so that children cannot access the unit Do not leave it unatte...

Страница 48: ...ST da 18 C a 38 C T da 18 C a 43 C da 10 C a 25 Manufacturer name and address Notified body identification number Year of production Ambient temperature class Gas refrigerant type Light power Defrost...

Страница 49: ...on Cut off power before removing the guard Earthing symbol Flammability Circuit containing flammable refrigerant Danger of heat High temperature parts Periodic condenser cleaning Glass door cleaning f...

Страница 50: ...s Table 11 Consumable materials Type Characteristics Application Image BATTERY 3V Lithium Coin Cell Battery type CR 2032 Clock battery on the board BATTERIA type AA rechargeable NiMH 1 2 V 2 7 Ah Acou...

Страница 51: ...quantity and leave space between the shelves and walls Place products in the cabinet a few at a time after the temperature has stabilised Material was placed in the freezer area at room temperature i...

Страница 52: ...External electromagnetic fields Spikes surge Icons do not react to commands or perform different functions Improper use of screen with pens and pins screen marked or spoiled Use stylus for touch scre...

Страница 53: ...coustic indication of a running alarm or failure touch the red triangle to get more information DMLP FAILURE RUNNING ALARM OR OFF if present ECT F FAILURE RUNING ALARM OR OFF TEMPERATURE ALARM OFF RUN...

Страница 54: ...sed Air condition the environment Check the electrical system or accidental disconnection of the plug LOW TEMPERATURE Contact Technical Support OPEN DOOR ALARM Check that door is properly closed Check...

Страница 55: ...d plug it again Switch on the equipment and touch the running failure alarm icon to eliminate the signalization Contact Technical Support MAINS FAILURE Check the electrical system or accidental discon...

Страница 56: ...alled Switch off the equipment seeSwitching Off Disconnect the plug from the mains wait for some seconds and plug it again Switch on the equipment and touch the running failure alarm icon to eliminate...

Страница 57: ...rsonnel who performed it 2 Have the problems been verified according to the information stated in the use and maintenance manual YES NO Fill out the following table Notes Note failure to submit a prop...

Страница 58: ...installed and another must be signed by the person in charge of the hospital ward and sent back to C F di Ciro Fiocchetti C s n c at one of the following addresses e mail tecnico fiocchetti it Fax 39...

Страница 59: ...7 The provided documentation is clear and it provides all the useful information for the usage of the device 1 2 3 4 5 8 The instructions stick on the packaging for its removal are clear and comprehen...

Страница 60: ...56 Annexes Medical Devices Touch Controller The following documents are attached Declaration of conformity with DIRECTIVE 93 42 ECC Electric safety check receipt Wiring diagram...

Страница 61: ...thermofisher com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE marcato col...

Страница 62: ...Italy 39 02 02 95059 434 254 375 Netherlands 31 76 571 4440 New Zealand 64 9 980 6700 Nordic Baltic CIS countries 358 9 329 100 Russia 7 812 703 42 15 Spain Portugal 34 93 223 09 18 Switzerland 41 44...

Отзывы: