1
TriPlus 300 Headspace Site Preparation
Entrance Requirements
2
TriPlus 300 Headspace Preinstallation Requirements Guide
Thermo Scientific
Entrance Requirements
Use the following guidelines to make sure the entrance to your site will allow delivery of the
TriPlus 300 Headspace sampling system.
1. Ensure the width of your delivery door opening is at least 80 cm (32 in.).
2. Make sure you have enough room to move boxes around corners, into elevators, or
through doorways. The table below contains the dimensions and weight of shipping
boxes, so that you can make accommodations:
Space and Load Requirements
TriPlus 300 Headspace sampling system must be installed on the left of the GC-GC/MS
units. Use the following tables to verify and determine the space and mass requirements for
the instrument of your GC-GC/MS system.
Table 1.
Container Dimensions
Container
Depth
Width
Height
(*)
Mass
(*)
TriPlus 300 Headspace Box 1:2
78.5 cm
(31 in.)
63.5 cm
(25 in.)
99.6 cm
(39 in.)
25 kg
(55 lbs)
TriPlus 300 Headspace Box 2:2
76.5 cm
(30 in.)
71.5 cm
(28 in.)
84 cm
(33 in.)
70 kg
(154 lbs)
(*) Including pallet