11
Upgrade Equipment
Upgrading a Stand Alone TRACE 1300/TRACE 1310 to MS Version
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
571
c.
Plug the power cable to the AC Input connector on the front of the module and to
the wall outlet. The LED marked
On
lights.
9. Close the front door of the GC.
10. If other external modules are present, plug the power cable to the AC Input connector of
each external module, and to the wall outlet.
11. Power on the GC.
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker), located at the back of the instrument, to the
position I.
12. Tune the ITQ and set its working conditions. Set the GC working conditions
accordingly.
For details please refer to the
TRACE 1300 and TRACE 1310 User Guide
and to the
ITQ
User and Hardware manuals
.
Coupling with the TSQ Quantum Mass Spectrometer
To configure the GC for the coupling with the TSQ Quantum mass spectrometer
1. Make sure that the preliminary operations have been carried out. See
2. Remove the right side panel.
a.
Open the front door of the GC.
b. Using a T20 Torxhead screwdriver, unscrew the left side panel screw from the interior
front panel. See
. Save the screw because it will be reused later.
Figure 631.
Right Side Panel Fixing Screw
c.
Slide the panel towards the back of the instrument up to the stop.