Upgrade Equipment
Upgrading a Stand Alone TRACE 1300/TRACE 1310 to MS Version
564
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
f.
On the left side wall in the interior of the oven, remove the partial cut plate from the
corresponding duct. See
Figure 620.
Perform the Duct for DSQ II (4)
4. Replace the left side panel with the left panel for MS provided.
a.
Remove the partial cut plate on the GC left panel for MS. See
Figure 621.
Left Panel for MS
b. Plug the ground wire previously removed to the left panel for MS provided.
c.
Place the left panel for MS and attach the screw holding it in place.
5. Introduce the DSQ II transfer line inner tube into the oven through the duct provided.
6. Attach the transfer line to the GC column using the proper nut and ferrule.
7. Place the Aux Temperature/Cryo module into the housing on the back of the GC.
See the section
“Adding an Aux Temperature/Cryo Module”
for details.
Note
T
he column must be conditioned before installing into the transfer line. See
“Installing the Column the First Time”
.
Duct for DSQ II Partial Cut Plate
Duct for DSQ II Partial Cut Plate