Upgrading a Stand Alone TRACE 1300/TRACE 1310 to MS Version
560
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 613.
Perform the Duct for ISQ Series / TSQ 8000 Series (4)
4. Replace the left side panel with the left panel for MS provided.
a.
Remove the partial cut plate on the GC left panel for MS. See
Figure 614.
Left Panel for MS
b. Plug the ground wire previously removed to the left panel for MS provided.
c.
Place the left panel for MS and attach the screw holding it in place.
5. Introduce the ISQ Series / TSQ 8000 Series transfer line inner tube into the oven
through the duct provided.
6. Attach the transfer line to the GC column using the proper nut and ferrule.
7. Close the front door of the GC.
8. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
9. Power on the GC.
Note
T
he column must be conditioned before installing into the transfer line. See
“Installing the Column the First Time”
Duct for ISQ Series and
TS 8000 Series Partial Cut Plate
Duct for ISQ Series and
TS 8000 Series Partial Cut Plate