Adding Modules
Adding a Generic Detector Interface
480
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
4. Reconnect the ground wire to the back cover terminal.
5. Replace the back cover of the GC proceeding in the reverse order in which it was
removed.
The result of the installation is shown in
Figure 508.
GDI Electrical Interface Installed into the GC
ATTENTION
TRACE 13100/1310 coupled with a TRACE 1310 Auxiliary Oven
If you are installing the GDI mechanical module in the
Front
or
Back
site of a TRACE
1300/1310 GC coupled with a TRACE 1310 Auxiliary Oven, the detector cable of the
GDI electrical interface must be connected as follows:
•
Mechanical module installed in the Front site
: On the backplane board disconnect
the flat cable from the connector marked
J13 DET.FRONT
. Connect the detector
cable to the connector
J13 DET.FRONT
.
•
Mechanical module installed in the Back site
:
Do NOT DISCONNECT the flat
cable from connector marked
J5 DET.REAR
. Connect the detector cable directly to
the additional connector on the flat cable.
Note
Continue the procedure of installation following the instruction reported in the
section
“Installing a GDI Mechanical Module”
.
GDI electrical interface