9
Adding Modules
Adding a PDD Module
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
459
Installing a PDD Module
To install a PDD detector module
The result of the operation is shown in
.
Figure 477.
Result of the PDD Installation
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
3. Close the gas supplies.
4. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
6. Remove the dummy module from the position where the detector module will be
installed.
a.
Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew and remove the two captive fixing
screws.
Note
The installation of a single PDD detector is considered. The injector modules are
deliberately missing for convenience.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.