7
Detectors Advanced Maintenance
Cleaning or Replacing the FPD Flame-side Heat Shields
314
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 309.
Filter Support Reinstallation
11. Reassembly the filter and the photomultiplier assembly.
a.
Reinsert the filter into the support. The mirror face must be oriented towards the
flame. See
.
Figure 310.
Filter Remounting
b. Reassembly the photomultiplier assembly and the detector body, then fix them
together tightening the knurled nut. See
CAUTION
Avoid touching the filter with your fingers. If you see fingertips on the filter,
clean it using a clean paper towel and, if needed, GC-grade methanol before remounting.
CAUTION
If the filter installed is equipped with the spacer (see the kits PN 19050785 and
PN 19050786), first insert the filter, and then the spacer.
Interferential Filter
Spacer