Thermo Scientific Thermo Scientific TDE Series Скачать руководство пользователя страница 7

2  |   Safety Considerations

Thermo Fisher Scientific ULT Freezers

Safety Considerations

In this manual, the following symbols and conventions are 
used:

Below are symbols and safety warnings that may be used on 
the product: (see list on pages 5 & 6)

Below are important safety precautions that apply to this 
product:

This symbol used alone indicates important 
operating instructions which reduce the risk of 
injury or poor performance of the unit.

CAUTION: 

This symbol, in the context of a 

CAUTION, indicates a potentially hazardous 
situation which if not avoided could result in 
minor to moderate injury or damage to the 
equipment.

WARNING: 

This symbol indicates potentially 

hazardous situations which, if not avoided, could 
result in serious injury or death.

WARNING: 

This symbol indicates situations 

where dangerous voltages exist and potential for 
electrical shock is present.

The snowflake symbol indicates extreme low 
temperatures and high risk of frostbite. Do not 
touch bare metal or samples with unprotected 
body parts.

This symbol indicates a need to use gloves during 
the indicated procedures. If performing 
decontamination procedures, use chemically 
resistant gloves. Use insulated gloves for handling 
samples and when using liquid nitrogen.

Before installing, using or maintaining this 
product, please be sure to read this manual and 
product warning labels carefully. Failure to follow 
these instructions may cause this product to 
malfunction, which could result in injury or 
damage.

Use this product only in the way described in the 
product literature and in this manual. Before using 
it, verify that this product is suitable for its 
intended use. If this equipment is used in a 
manner not specified by the manufacturer, the 
protection provided by the equipment may be 
impaired.

Do not modify system components, especially 
the controller. Use OEM exact replacement 
equipment or parts. Before use, confirm that the 
product has not been altered in any way.

WARNING: 

Your unit must be properly grounded 

in conformity with national and local electrical 
codes. Never connect the unit to overloaded 
power sources.

WARNING: 

Disconnect the unit from all power 

sources before cleaning, troubleshooting, or 
performing other maintenance on the product or 
its controls.

WARNING: 

“Caution, risk of fire”. This unit is 

charged with hydrocarbon refrigerants.

WARNING: 

 Ensure all ventilation openings are 

not obstructed.

WARNING: 

 Do not use mechanical devices or 

other means to accelerate the defrosting 
process, other than those recommended by the 
manufacturer.

WARNING: 

 Do not damage the refrigerant 

circuit.

WARNING: 

 In order to reduce flammability 

HAZARDS the installation of this equipment shall 
only be carried out by a suitably qualified person.

WARNING: 

 Use of this equipment adjacent to or 

stacked with other equipment should be avoided 
because it could result in improper operation. If 
such use is necessary, this equipment and the 
other equipment should be observed to verify 
that they are operating normally. 

WARNING: 

Use of accessories, transducers and 

cables other than those specified or provided by 
the manufacturer of this equipment could result in 
increased electromagnetic emissions or 
decreased electromagnetic immunity of this 
equipment and result in improper operation.

Содержание Thermo Scientific TDE Series

Страница 1: ...Ultra Low Temperature Freezers Thermo ScientificTM TDE Series Installation and Operation 332851H01 Revision A July 2022...

Страница 2: ...e performed by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scien...

Страница 3: ...onitoring system 8 General Usage 8 Initial Loading 8 Grille Door Opening Closing 8 Operating Standards 9 Electrical Specifications 9 Installation 10 Location 10 Protective Conductor current 10 Wiring...

Страница 4: ...t Up 17 Operation 17 Chart Recorders Optional 18 Set Up and Operation 18 Changing Chart Paper 18 Calibration Adjustment 18 Maintenance 19 Cleaning the Condenser 19 Cleaning the Condenser Filter 19 Gas...

Страница 5: ...Appendix A Alarm Summary 28 Appendix B Modbus ASCII Parameter Table 30 End of Life Care 38 Contact Information 39...

Страница 6: ...Thermo Fisher Scientific ULT Freezers Models 1 Models Table 1 Applicable Models Model Size xxx Voltage Description TDECxxx86L 063 252 396 A D V Medical Device...

Страница 7: ...the product literature and in this manual Before using it verify that this product is suitable for its intended use If this equipment is used in a manner not specified by the manufacturer the protecti...

Страница 8: ...rigerant used To be repaired only by trained service personnel Do not puncture refrigerant tubing AVERTISSEMENT Risque d incendie ou d explosion R frig rant inflammable utilis Pour tre r par uniquemen...

Страница 9: ...rs are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 1 kV for input output lines 2 kV for AC powerlines 1 kV for input outpu...

Страница 10: ...tes uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipme...

Страница 11: ...ied hazardous locations nor to be used for the storage of flammable inventory Symbols Glossary NOTICE AVIS WE RECOMMEND THE USE OF A REDUNDANT AND INDEPENDENT TEMPERATURE MONITORING SYSTEM AND THAT FR...

Страница 12: ...authorization When submitting a claim for shipping damage request that the carrier inspect the shipping container and equipment The packaging can be stored and re used Packing List Inside the freezer...

Страница 13: ...point before loading the next shelf Repeat this process until the freezer is fully loaded Fill a chest by starting at the left side near the probe Filling with room temperature racks will result in a...

Страница 14: ...30 V 60 Hz Main supply voltage fluctuations not to exceed 5 or 10 of the nominal voltage for 100 V 60 Hz Electrical Specifications The last character in the model number listed on the data plate ident...

Страница 15: ...ed on the lower left side of the unit WARNING Use care when closing the door lid so that fingers are not pinched between cabinet and door lid CAUTION Connect the equipment to the correct power source...

Страница 16: ...the wiring from the remote alarm to the connector install the connector to the freezer micro board The pin configuration is shown in Figure 3 below Figure 3 Remote Alarm Pin Configuration For systems...

Страница 17: ...m a warm start condition Cold Alarm value The range is 99 C to within 5 C of setpoint Note A setpoint change may automatically change the warm cold alarm setpoints as well to maintain a minimum 5 C se...

Страница 18: ...lay Warm Alarm Test Press the plus and checkmark buttons at the same time to initiate the warm alarm test During the warm alarm test the actual cabinet temperature will not be displayed The display te...

Страница 19: ...are showing Press the checkmark while YES is showing With a flashing YES you are required to confirm by pressing the checkmark button a second time Power down will then be complete If the checkmark bu...

Страница 20: ...ey provide useful safety information about installation use and maintenance to help avoid mishaps and possible accidents Remove the packing and make sure the scraper and magnetic hook are in good cond...

Страница 21: ...liquid from the bottom of the cylinder CO2 cylinders must be kept at room temperature to function properly LN2 bottles are functional at any reasonable temperature WARNING If a CO2 or LN2 cylinder fa...

Страница 22: ...e flexible hose to sharp bends Doing so may shorten the life of the hose Start Up When the unit is started it will recognize if a backup system is installed 1 Follow the instructions in Backup System...

Страница 23: ...d The pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the left si...

Страница 24: ...off the freezer 3 Open the outer door lid and all inner doors sublids 4 Let the freezer stand with doors lid open for at least 24 hours This allows both the interior and foamed refrigeration system t...

Страница 25: ...edule Regular maintenance is important to keep the unit working properly Inspect clean as directed in the manual Item Interval Ice Frost build up To maintain proper closure of sublids remove any ice o...

Страница 26: ...on with a minimum of 8 20 cm of space on the top and sides 6 15 cm in back Icing Frost due to high relative humidity Check if the location meets requirements Maximum relative humidity 60 for temperatu...

Страница 27: ...nit must be properly grounded in conformity with national and local electrical codes Troubleshooting procedures involving live voltage is dangerous and if done improperly can result in injury and or d...

Страница 28: ...t is constantly alarming Exterior lid is closed but not sealed completely Clean any ice build up on gasket and or cabinet surface Check for punctures or tears on gasket Door open alarm exterior door n...

Страница 29: ...noise Unit is not level Check if the unit is installed in a level area free from vibration Refer to Leveling in user manual Loose side panels Check side panel screws tighten them if necessary Rubber t...

Страница 30: ...ure E03 Heat Exchanger Probe Failure E04 Power Failure E05 Failure to Reach Setpoint E06 BUS Battery Low Voltage E07 System Battery Low Voltage E08 Lost Communication Failure Main to UI E09 Lost Commu...

Страница 31: ...Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original five year warranty period T...

Страница 32: ...lights light bulbs batteries and door gaskets are excluded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equi...

Страница 33: ...erature variation will increase Contact customer service Display will intermittently show E03 Ambient Probe Failure Wrench Yes Ambient Probe TC has malfunctioned This doesn t affect the performance of...

Страница 34: ...Battery voltage is too low If error persists the battery may need to be replaced Display will intermittently show E06 Stuck Button Alarm Wrench Yes A button has been pressed for more than 5 minutes D...

Страница 35: ...ata Data Type Remarks 1 cabinet setpoint c 0x03 530 2 3a 31 39 30 33 30 35 33 30 30 30 30 32 41 44 0d 0a 3A 31 39 30 33 30 34 46 46 46 46 46 43 45 30 30 32 0D 0A 0xFFF FFCE 0 int Convert the data valu...

Страница 36: ...0 33 31 35 32 32 30 32 30 32 30 32 30 32 30 32 30 32 30 32 30 32 30 32 30 44 30 0D 0A 155D R0A0 1R String All the remaining are spaces 5 Product ID 0x03 598 2 3a 31 39 30 33 30 35 39 38 30 30 30 32 34...

Страница 37: ...www h schmidt net FloatConverter IEEE754 html 2 Paste the data in Hexadecimal Representation and press Enter 3 The temp value is shown in Decimal representation 9 TC2 0x03 4CC 2 3a 31 39 30 33 30 34...

Страница 38: ...with below steps 1 Open the link https www h schmidt net FloatConverter IEEE754 html 2 Paste the data in Hexadecimal Representation and press Enter 3 The temp value is shown in Decimal representation...

Страница 39: ...resentation and press Enter 3 The temp value is shown in Decimal representation 16 TC9 0x03 4E8 2 3a 31 39 30 33 30 34 45 38 30 30 30 32 46 36 0d 0a 3A 31 39 30 33 30 34 34 31 42 45 32 34 41 31 31 34...

Страница 40: ...D Temperature value is 78 C 19 Alarms 0x03 514 2 3a 31 39 30 33 30 35 31 34 30 30 30 32 43 39 0d 0a 3A 31 39 30 33 30 34 30 30 30 30 30 34 30 32 44 36 0D 0A 0x00 0004 02 unit 1 Active 0 Inactive b19 B...

Страница 41: ...Bus Solenoid Injection b1 Bus Pressure switch b2 Reserved for factory use only b3 Line voltage circuit state change normal buck boost b4 Compensated line voltage change b5 Reserved for factory use onl...

Страница 42: ...30 30 30 32 41 39 0d 0a 3A 31 39 30 33 30 34 30 30 30 30 30 30 30 30 44 43 0D 0A 0xFF FFFC E0 int Convert the data value into signed 2 s complement and divide with 10 which gives the setpoint Ex Signe...

Страница 43: ...ed household waste Please use the public collection system to return recycle or treat them in compliance with the local regulations Remove all samples and items before defrosting a unit to room ambien...

Страница 44: ...This page is intentionally left blank...

Страница 45: ...This page is intentionally left blank...

Страница 46: ...ntact information for your area Countries Sales North America 1 866 984 3766 India toll free 1800 22 8374 India 91 22 6716 2200 China 800 810 5118 400 650 5118 Japan 81 3 5826 1616 France 33 2 2803 21...

Отзывы: