![Thermo Scientific Thermo Scientific TDE Series Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/thermo-scientific/thermo-scientific-tde-series/thermo-scientific-tde-series_manual_1104666034.webp)
Thermo Fisher Scientific ULT Freezers
Appendix A: Alarm Summary | 29
Wrong Model Alarm
Wrench
Yes
Invalid Control Model Alarm. Contact service to ensure the
correct model is selected for the system to avoid cargo
loss.
Display will intermittently show “E00”.
Firmware Build
Incompatible
Wrench
Yes
Firmware build indicates incompatibility that can result in
modules to be non-coherent.
Display will intermittently show “E01”.
System Battery
Low Voltage Alarm
Wrench
Yes
System Battery voltage is too low. If error persists, the
battery may need to be replaced.
Display will intermittently show “E07”.
BUS Battery
Low Voltage Alarm
Wrench
Yes
BUS Battery voltage is too low. If error persists, the battery
may need to be replaced.
Display will intermittently show “E06”.
Stuck Button Alarm
Wrench
Yes
A button has been pressed for more than 5 minutes.
Display will intermittently show “E10”.
System Battery
Failure Alarm
Wrench
Yes
System Battery disconnected or failed.
Display will intermittently show “E12”.
BUS Probe
Failure Alarm
Wrench
Yes
BUS cannot detect temperature. BUS will continuously
inject. Contact Customer service.
Display will intermittently show “E13”
BUS Battery
Failure Alarm
Wrench
Yes
BUS battery disconnected or failed.
Display will intermittently show “E14”.
Wrong Power Alarm
Wrench
Yes
The unit has detected the Wrong Power connected. Verify
the proper voltage.
Display will intermittently show “EA1”.
Alarm Summary
Alarm Message
Warning
Icon
Remote
Alarm Event
Description
Содержание Thermo Scientific TDE Series
Страница 44: ...This page is intentionally left blank...
Страница 45: ...This page is intentionally left blank...