Switch A should be “OFF”.
Before starting the Auto Run, specify, a run time as follows:
Set switch B to “RUN”. Adjust switch C until the display shows the desired duration
of the run in hours and minutes.
Start the run by setting switch A to “AUTO”.
During the run, you can toggle the unit to display the following information:
•
Remaining run time. Set switch B to “RUN”. The display now
counts downward, showing the hours and minutes of time left
in the current run.
or
•
Remaining heater time. Set switch B to “HEATER”. If a specific
time setting is selected for the HEATER, then the display counts
down and indicates the amount of time remaining before the
HEATER shuts off.
To terminate a run, set switch A to “OFF”.
3.4
EXAMPLE OF AN AUTO RUN
Thermal energy can be applied to the concentrator chamber to counteract the
cooling effect of evaporation, it maintains samples in the liquid state, and
accelerates the concentration run. The DRYING RATE switch selects the desired
drying rate:
“LOW”
maintains the concentrator chamber at ambient temperature.
“MEDIUM”
maintains the concentrator chamber at 43° C.
“HIGH”
maintains the concentrator chamber at 65° C.
Select the highest drying rate if you desire accelerated drying. To ensure against
overdrying of samples, and possible denaturing, it is recommended that the
HEATER timer be set for only a portion of the total run time (e.g., 50%-75% of total
run time).
3.5
DRYING RATE
6