Temperature Probe Calibration
This section applies only to the calibration of the temperature probe system. The hot plate surface and the motor speed
cannot be calibrated. This method is only applicable to units that have heating functionality.
Enter the System Configuration by pressing and holding the knob. After about 5 seconds a single you may remove your
finger from the knob. The display indicate “system config” on the first row, the “Temp Unit” will show on the second row
of the display. Turn the knob clockwise until the “Calibrate Temp” show on the second row of the display. If you do no
action, the display will return to the operation status. The option “Calibrate Temp” will only show when there is a remote
probe PT100 or PT1000 adopted.
Press the knob to enter the Calibrate Temp setting interface. The “Calibrate Temp” show on the first row, the measured
temperature flash on the second row. If the data is not matched the temperature you measure by external thermo
detector, you can turn the knob to calibrate. When it’s matched, press the knob to complete the calibration.
When setting and/or viewing is complete, the knob can be pressed to return to the previous operation menu.
Содержание SUPER-NUOVA+ HP88850190
Страница 16: ...16 300 540 C 1 2...
Страница 17: ...17 1 2 3 4 5 6 CIC0000793 CIC0000469 7 8 9 35 15 9 10 10 25 11 3 7 x 7 10 50 C 11 12 13 14...
Страница 36: ...36 7x7 Hot Plate Exploded View Exploded Views...
Страница 37: ...37 EXPLODED VIEWS 7x7 Hot Plate Top Assembly...
Страница 38: ...38 EXPLODED VIEWS 10x10 Hot Plate Exploded...
Страница 39: ...39 EXPLODED VIEWS 10x10 Hot Plate Top Assembly...