9
Maintenance
Replacing the gas inlet filter
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
9-9
6. Turn the airbox upside down, unlatch the tabs (5/
)on the left-hand side from the HEPA
filter tabs (6/
).
7. Unlatch the tabs on the right-hand side (3/
) of the airbox (1/
) from the
corresponding slots provided in the HEPA-Filter (4/
).
Figure 9-7.
Installing the HEPA-filter
8. Install the new HEPA-filter in airbox (1/
) and engage the locking tabs.
9. Place the airbox on its seat in the water reservoir cover panel.
10. If the unit has been operated without the HEPA-filter previously, activate the HEPA-filter in the
user configuration according to instructions
“Activating / de-activating the HEPA-filter:“
.
11. If necessary, set the reminder interval for the HEPA-filter replacement in the user configuration
according to instructions
. The reminder interval for
HEPA-filter replacement may be set between 1 and 12 months. We recommend setting this
parameter to the factory preset value of 6 months.
Replacing the gas inlet filter
The gas inlet filter (CO
2
-/O
2
-/N
2
supply) has a plastic thread and is screwed by hand into the female
thread socket at the control box.
CLICK
1
2
3
3
4
4
7
8
5
6
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...