Some of the features of the SPECTRONIC 200 spectrophotometer are:
• CCD polychromator based spectrophotometer
• Large, easily accessible sample compartment
• Removable sample compartment for easy cleaning
• Adjustable display angle
• 320 x 240 pixel color graphical TFT LCD display
• PCL 5E and PCL 6 printer compatibility
• Instrument can be stored on a shelf in a vertical position
• No moving parts to wear out or go out of alignment
• Convenient sample holders for cuvettes and test tubes
• Built-in cuvette rack wells
Technical Specifications
• Spectral Bandwidth:
≤ 4 nm
• Light Source:
Tungsten-halogen lamp
• Detector:
CCD
• Display:
Color Graphical LCD
• Scan Ordinate Modes:
%T, Abs
• Wavelength range:
340-1000 nm
• Monochromator:
600 lines/mm holographic grating
• Wavelength readability:
1 nm on board, 0.5 nm under computer control
• Wavelength Accuracy:
±2 nm
• Wavelength
reproducibility:
± 1 nm
• Stray light:
< 0.2 %T (with SPECTRONIC standard SRM 400 filter)
• Photometric range:
-0.3A to 2.5A
• Photometric Accuracy:
± 0.01A at 0.3A
± 0.05A at 1.0A
(SPECTRONIC standard filter measured at 590 nm)