3-19
Graphical User Interface
Figure 3–35: Status – Alarm Screen: Operating Details Expanded
NOTICE
These status information fields show the same content as the “Status” screen, plus the rotor log.
This is described in the sections
, respectively.
Proceed as follows to view the alarm list with alarm details:
1. If you wish to silence the audible alarm (and hide the alarm message) temporarily, tap the Snooze button in
the “Info & Health Status” area of the touchscreen (see
for an example)
.
2. Tap on the alarm bell icon in the “Info & Health Status” area.
The “Status - Alarm” screen shown in
appears, with the latest alarm list item already expanded.
3. To view background information about the centrifuge and rotor, tap the field header to expand the “Operating
, “Rotor-Centrifuge Lifecycle”, “Rotor Log”, or “Imbalance Alarms” field.
4. Tap the field header again to hide the content of the information field.
5. Read the description and follow the instructions to correct the problem—for example:
a. Return to the Home screen.
b. Press the Lid Open button .
c. Remove the rotor and balance the load of the buckets.
d. Reinstall the rotor and restart the centrifuge.
6. When you have corrected the problem, tap the
Acknowledge
button to acknowledge and clear the alarm.
The checkbox next to the alarm list entry is activated.
NOTICE
If you acknowledge the alarm without correcting the problem, the alarm will return immediately.
7. If you wish to view more alarms from the list, tap the scrollbar and drag to scroll down.
8. Tap the desired alarm list item.
The alarm item is expanded to reveal the details.
9. Solve the problem and acknowledge the alarm, then tap the alarm list item one more time to reduce it.
Once you have solved and acknowledged all alarms, the “Status – Good” Screen appears to confirm that
the centrifuge is free of alarms, as shown in
below.
Figure 3–36: Status – Screen
10. Tap the
X
icon in the top right corner to return to the “Home” screen.
Содержание Sorvall ST Plus Series
Страница 1: ...Thermo Scientific Sorvall X Pro ST Plus Series Centrifuges Instructions for Use 50158527 d 08 2019 ...
Страница 116: ...B 9 TX 750 B 1 4 Biocontainment Certificate ...
Страница 120: ...B 13 TX 1000 B 2 4 Biocontainment Certificate ...
Страница 126: ...B 19 HIGHConic II B 4 4 Biocontainment Certificate ...
Страница 133: ...B 26 Fiberlite F15 6 x 100y B 7 4 Biocontainment Certificate ...
Страница 137: ...B 30 HIGHPlate 6000 B 9 4 Biocontainment Certificate ...
Страница 140: ...B 33 M 20 Microplate B 10 4 Biocontainment Certificate ...
Страница 143: ...B 36 Microliter 48 x 2 B 11 4 Biocontainment Certificate ...
Страница 146: ...B 39 Microliter 30 x 2 B 12 4 Biocontainment Certificate ...
Страница 149: ...B 42 MicroClick 30 x 2 B 13 4 Biocontainment Certificate ...
Страница 152: ...B 45 MicroClick 18 x 5 B 14 4 Biocontainment Certificate ...
Страница 155: ...B 48 Fiberlite F21 48 x 2 B 15 4 Biocontainment Certificate ...
Страница 159: ...B 52 Fiberlite F10 6 x 100 LEX B 17 4 Biocontainment Certificate ...