Troubleshooting
Troubleshooting by Guide
56
Sorvall ST 8 Centrifuge
Thermo Scientific
E-025;
E-027
Read Manual
Check, if the centrifuge lid is blocked.
Restart the centrifuge.
If the error message appears again, inform the customer service.
E-029;
E-045
Read Manual
Check, if a rotor is installed.
Check, if the rotor is qualified for the Sorvall ST 8 Centrifuge
(check
“Rotor Selection”
on
page 20
).
Restart the centrifuge.
If the error message appears again, inform the customer service.
E-030
Power Failure
Check the power supply of the centrifuge. Make sure not to
operate too many devices at one power source.
Let the centrifuge cool down for 15 minutes.
If the error message appears again, inform the customer service.
E-098
Imbalance Load
Check the load placed in the rotor.
Check that the rotor cross bolts are greased well.
Restart the centrifuge.
If the error message appears again, inform the customer service.
E-060
Temp Low!
CAUTION
Icy metal parts!
Restart the centrifuge.
If the error message appears again, inform the customer service.
E-046
Door Open!
Restart the centrifuge.
If the error message appears again, inform the customer service.
Error number
Error message
Troubleshooting
Содержание Sorvall ST 8
Страница 2: ......
Страница 3: ...Thermo Scientific Sorvall ST 8 Centrifuge Instruction Manual 50138912 a February 2013 ...
Страница 13: ...Preface 14 Sorvall ST 8 Centrifuge Thermo Scientific ...
Страница 25: ...2 Before Use Shipping the Centrifuge 26 Sorvall ST 8 Centrifuge Thermo Scientific ...
Страница 26: ...Thermo Scientific Sorvall ST 8 Centrifuge 27 3 Control Panel Contents Control Panel on page 28 Keys on page 28 ...
Страница 43: ...5 System Menu Description 44 Sorvall ST 8 Centrifuge Thermo Scientific ...
Страница 51: ...7 Disposal 52 Sorvall ST 8 Centrifuge Thermo Scientific ...
Страница 57: ...8 Troubleshooting When to Contact Customer Service 58 Sorvall ST 8 Centrifuge Thermo Scientific ...
Страница 65: ...10 Contact 66 Sorvall ST 8 Centrifuge Thermo Scientific ...
Страница 67: ......