4
Operation
Centrifugation
4-10
Legend X1 / X1R
Thermo Scientific
Overwriting a passcode-protected program
If you wish to overwrite a passcode-protected program proceed as follows:
1. Select the program you wish to change.
2. Press the program memory key to load any of the programs 1-5.
Press the DIRECTORY key
in order to load any of the programs 1-99 by means of the
numeric pad.
Enter the program number and confirm with ENTER
.
3. Enter the passcode.
If you enter the wrong password, the following message appears:
The display restores the current values. Return to step 1.
Centrifugation
Once the rotor has been properly installed, the main switch has been turned on and the lid has been
closed, you can start centrifuging.
Starting centrifuge program
Press the START key
on the control panel. The centrifuge accelerates to the pre-set speed with
the time display active.
If the speed setting is higher than the maximum permissible speed or RCF-value for the particular
rotor, then the display will show the message CHECK SET SPEED. The centrifuge decelerates.
Please enter a new value.
You cannot open the lid as long as the centrifuge is running.
Imbalance indicator
If a load is imbalanced, this will be indicated at speed higher than approx. 300 rpm by the message
"Imbalanced load".
The run will terminate.
Check the loading and start the centrifuge once again. See the information on proper loading in the
rotor instruction manual. For information on troubleshooting, see section
“Troubleshooting by
User”
on
page 7-3
.