The Main Screen
8 of 73
3-
Accumulated Centrifugal Effect (ACE)
Input as X, Y * 10
z
X: Pre-decimal digits
Y: Decimal places
Z: Exponent of power of ten
1.2.2
Speed
The rotation speed can be entered either in revolutions per minute (rpm) or as relative centrifugal force (RCF).
Acceleration cannot continue beyond the maximum speed of the rotor currently in use.
1.2.3
Temperature
The temperature is entered in degrees Celsius (°C). Positive and negative values may be entered within the range
from -20 °C to +40 °C. If this value is changed, the centrifuge cools the rotor chamber to the specified temperature
straight away.