10 | Technical Specifications
Smart2Pure Pro Ultrapure water system
Technical Specifications
Note:
Check at regular intervals the quality of your feed-water.
*RO flows at 15°C
Feed-water requirements
Source
Potable tap water pretreated with Barnstead Pretreatment System (50157886)
Silt density index (SDI)
<3
Turbidity
<1 NTU
Conductivity (reference
tempe25 °C)
1500 μS/cm (670 Ω·cm)
Hardness
Hardness stabilized
Chlorine
Max. 0.1 ppm
pH range
4 - 11
Temperature
+2°C - +35°C
Pressure
2 - 6 bar / 29 - 87 PSI (at a pressure >6 bar / 87 PSI a pressure reducer must be
installed upstream of the system.
Product water quality ASTM Type I
(from dispensing valve)
UV/UF
Conductivity (reference
tempe25°C)
μS/cm
0.055
Resistance (reference
tempe25°C)
MΩ·cm
18.2
Product water quality ASTM Type II
(tank quality)
UV/UF
Conductivity (reference
tempe25°C)
μS/cm
0.1 – 1.0
Resistance (reference
tempe25°C)
MΩ·cm
1 - 10
Retention rate for bacteria and
particles
%
99
External tank capacity
Smart2Pure Pro
Liter
30 or 60
Smart2Pure Pro RO cartridge
permeate flow*
L/hr
16lph
+/-15%
Smart2Pure Pro RO cartridge
concentrate flow*
L/hr
60
TOC value
ppb
1 - 5
RNase
ng/ml
<0.003
DNase
pg/ul
<0.4
Bacteria
CFU/ml
<0.
0
1
Bacterial endotoxins
EU/ml
<0.001
Particle >0.2μm, #/ml
μm/ml
<1
Dispense rate
L/min
up to 1