Additional configuration
Thermo Fisher Scientific
Smart-Vue 4-20 mA Module Installation Guide
17
Additional
configuration
Your Smart-Vue module offers many features that are controlled by your
Smart-Vue Server/Smart-Vue Client software. Use the software to define
settings for measurement period, frequency of data transmission, alarm
limits for sending alerts via phone, fax, e-mail and SMS, sensor
calibration/correction parameters, archiving and more. Enter your login
name and password when prompted.
You may download sensor correction/calibration parameters directly over
the Internet if your purchase included calibration services. Please see the
Thermo Scientific Smart-Vue Software User Manual
for more details.
Setting basic sensor parameters
1. Access end-point module settings by pressing F11 or selecting
Settings ==> Sensor settings (
) on the Smart-Vue Client main
screen.
Figure 11.
Main sensor settings screen
2. Choose a sensor in the left-hand panel, and adjust the specific
settings (Sensor Settings tab). You should do this for all sensors in
your system.
3. Click on the plus sign(
+)
next to the Unit field to assign your own
specific unit values to the measured current values. For example, for
a device output of -100°C to +200°C the 4mA value would equal
-100 and the 20mA value would equal +200°C. Remember to insert
the negative symbol, if applicable. Click on OK.
Click on dials
(dots) or
double-click
Hours/Minutes
to set