![Thermo Scientific SHKA4450 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/thermo-scientific/shka4450/shka4450_manual_1104780010.webp)
10
G
ENERAL
S
PECIFICATIONS
Catalog Number
RS232 Interface*
Recorder Output*
(Located on left side
rear of shaker)
Optional Platform
Dimensions L X W
Exterior Dimensions
L X W X H
SHKA4450
SHKA4450CC
None
None
13" X 11"
(33.0 X 27.9 cm)
27.16" X 14.09" X
15.75"
(69.0 X 35.8 X
40.0 cm)
SHKA4450-1CE
SHKA4450CC-1CE
None
None
13" X 11"
(33.0 X 27.9 cm)
27.16" X 14.09" X
15.75"
(69.0 X 35.8 X
40.0 cm)
SHKE4450
SHKE4450CC
Monitor speed, tem-
perature and time
with a computer
10 mv/C output mon-
itors temperature
with external chart
recorder
13" X 11"
(33.0 X 27.9 cm)
27.16" X 14.09" X
15.75"
(69.0 X 35.8 X
40.0 cm)
SHKE4450-1CE
SHKE4450CC-1CE
Monitor speed, tem-
perature and time
with a computer
10 mv/C output mon-
itors temperature
with external chart
recorder
13" X 11"
(33.0 X 27.9 cm)
27.16" X 14.09" X
15.75"
(69.0 X 35.8 X
40.0 cm)
Pollution Degree**
2
Installation Category**
II
Altitude
2000 meters MSL (Mean Sea Level)
Relative Humidity
20% to 80% maximum, non-condensing
Electrical Supply
120 VAC or 240 VAC
Voltage Tolerance
±10% of normal rated line
Temperature
15ºC to 32ºC
Product Usage
This product is intended for use indoors only.
Unit's Environmental Operating Conditions
*Interface cables not to exceed 9.8' (3 m) in length.
**Refer to IEC 664-1
Caution
It is not recommended to operate shaker in a CO2 enriched atmosphere. The formation of carbonic
acid could cause electrical failures.