Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
109
8
Compliance Information
Declaration of Conformity
Thermo Fisher Scientific certifies this product was tested thoroughly, inspected, and found to meet its published specifications
when it was shipped from the factory.
The
DataStick
is a DC powered measurement system with many sensing and communications options. Representative
combinations of the product have been tested and are certified as indicated to the following instrumentation standards:
Product Safety:
Certified by Thermo Fisher Scientific LLC to EN 61010-1
Immunity:
The equipment was tested to the Industrial level of performance in accordance with EN 61326, which includes
test criteria for electromagnetic compatibility for electrical equipment for measurement control and laboratory use per
89/336/EEC EMC. Supporting test records were written and certified by Thermo Fisher Scientific LLC.
Test Standards include:
IEC 1000-4-2:1995 (EN61000-4-2:1995) Electrostatic Discharge Immunity Criteria B
IEC1000-4-3: 1995 (EN61000-4-3:1996) Radiated RF Electromagnetic Field Immunity Criteria A
IEC1000-4-4: 1995 (EN61000-4-4: 1995) Electrical Fast Transient/Burst Criteria B
IEC1000-4-5: 1995 (EN61000-4-5: 1995) Surge Criteria B
IEC1000-4-6: 1996 (EN61000-4-6: 1996) Conducted Disturbances Induced by RF fields Criteria A
Emissions:
The equipment was tested to for radio frequency emissions in accordance with EN 61326:1998, which
includes test criteria for electromagnetic emissions limits for electrical equipment for measurement, control and laboratory
use. Supporting test records were written and certified by Thermo Fisher Scientific LLC.
Test Standards include:
EN55011 (CISPER 11) Conducted emissions from 150 kHz to 30 MHz Class B.
EN55011 (CISPER 11) Radiated emissions from 30 MHz to 1 GHz Class B.